FUSION SARS-COV-2
Report
- Report Number
- 2024800-2022-01099
- Event Type
- Malfunction
- Date Received
- October 11, 2022
- Date of Event
- September 11, 2022
- Report Date
- October 11, 2022
- Manufacturer
- HOLOGIC, INC.
- Product Code
- OJR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
LOGS WERE REVIEWED BY HOLOGIC FIELD SERVICE ENGINEER (FSE) AND CONCLUDED THAT THE INCREASE IN POSITIVITY RATES APPEARED TO HAVE BEGUN JUST AFTER A SAMPLE SHIELD MAINTENANCE. HOLOGIC FIELD APPLICATIONS SERVICES (FAS) TESTED ENVIRONMENTAL SWABS FROM THE CUSTOMER¿S PANTHER FUSION INSTRUMENT AND SURROUNDING AREAS WITH THE FUSION SARS-COV-2 ASSAYS PRIOR TO PERFORMING A CLEANING. THE PRE-CLEANING ENVIRONMENTAL RUN RESULTED IN SEVERAL SARS-COV-2 POSITIVE SITES. FAS PERFORMED A THOROUGH DECONTAMINATION OF THE INSTRUMENT AND SURROUNDING AREAS AND COLLECTED ENVIRONMENTAL SWABS. FAS OBSERVED THAT THE SAMPLE SHIELD ON BOARD WAS WARPED REDUCING THE PIPETTOR CLEARANCE OF THE SHIELD RESULTING IN POTENTIAL CONTACT BETWEEN THE SHIELD AND THE SAMPLE TIP. FAS ALSO NOTED THAT ENVIRONMENTAL CONTAMINATION COULD POTENTIALLY BE CAUSED FROM NOT CHANGING GLOVES AND SAMPLES BEING IN CONTACT WITH CLEAN REAGENT PREP AREAS. THE POST-CLEANING ENVIRONMENTAL RUN RESULTED IN NO FURTHER INDICATION OF CONTAMINATION. HOLOGIC DETERMINED THAT PANTHER SN: 03517 IS NO LONGER CONTAMINATED. HOLOGIC ALSO RECOMMENDED REFRESHER AND PROFICIENCY TRAINING FOR THE PANTHER FUSION SYSTEM IN THE HERVEY BAY LABORATORY AND ALL PATHOLOGY QUEENSLAND LABORATORIES TO ENSURE THE EFFICACY OF THE SYSTEM AND OPERATOR EXPERTISE.
CUSTOMER CONTACTED HOLOGIC REGARDING AN INCREASED POSITIVE RATE AND POTENTIALLY DISCREPANT SARS-COV-2 RESULTS PRODUCED BY THE PANTHER FUSION SN (B)(4). CUSTOMER REPORTED THAT THERE WERE 7 ALLEGED FALSE POSITIVE RESULTS. CUSTOMER RETESTED USING A DIFFERENT PLATFORM (ALINITY & GENEXPERT). THE LOT NUMBER OF THE ASSAY USED WAS NOT PROVIDED TO HOLOGIC.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 390973 | FUSION SARS-COV-2 | REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID | OJR | HOLOGIC, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |