FDA Adverse Event Malfunction Summary report: N

FUSION SARS-COV-2

MDR report key: 15584035 · Received October 11, 2022

Report

Report Number
2024800-2022-01099
Event Type
Malfunction
Date Received
October 11, 2022
Date of Event
September 11, 2022
Report Date
October 11, 2022
Manufacturer
HOLOGIC, INC.
Product Code
OJR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

LOGS WERE REVIEWED BY HOLOGIC FIELD SERVICE ENGINEER (FSE) AND CONCLUDED THAT THE INCREASE IN POSITIVITY RATES APPEARED TO HAVE BEGUN JUST AFTER A SAMPLE SHIELD MAINTENANCE. HOLOGIC FIELD APPLICATIONS SERVICES (FAS) TESTED ENVIRONMENTAL SWABS FROM THE CUSTOMER¿S PANTHER FUSION INSTRUMENT AND SURROUNDING AREAS WITH THE FUSION SARS-COV-2 ASSAYS PRIOR TO PERFORMING A CLEANING. THE PRE-CLEANING ENVIRONMENTAL RUN RESULTED IN SEVERAL SARS-COV-2 POSITIVE SITES. FAS PERFORMED A THOROUGH DECONTAMINATION OF THE INSTRUMENT AND SURROUNDING AREAS AND COLLECTED ENVIRONMENTAL SWABS. FAS OBSERVED THAT THE SAMPLE SHIELD ON BOARD WAS WARPED REDUCING THE PIPETTOR CLEARANCE OF THE SHIELD RESULTING IN POTENTIAL CONTACT BETWEEN THE SHIELD AND THE SAMPLE TIP. FAS ALSO NOTED THAT ENVIRONMENTAL CONTAMINATION COULD POTENTIALLY BE CAUSED FROM NOT CHANGING GLOVES AND SAMPLES BEING IN CONTACT WITH CLEAN REAGENT PREP AREAS. THE POST-CLEANING ENVIRONMENTAL RUN RESULTED IN NO FURTHER INDICATION OF CONTAMINATION. HOLOGIC DETERMINED THAT PANTHER SN: 03517 IS NO LONGER CONTAMINATED. HOLOGIC ALSO RECOMMENDED REFRESHER AND PROFICIENCY TRAINING FOR THE PANTHER FUSION SYSTEM IN THE HERVEY BAY LABORATORY AND ALL PATHOLOGY QUEENSLAND LABORATORIES TO ENSURE THE EFFICACY OF THE SYSTEM AND OPERATOR EXPERTISE.

Description of Event or Problem · 0

CUSTOMER CONTACTED HOLOGIC REGARDING AN INCREASED POSITIVE RATE AND POTENTIALLY DISCREPANT SARS-COV-2 RESULTS PRODUCED BY THE PANTHER FUSION SN (B)(4). CUSTOMER REPORTED THAT THERE WERE 7 ALLEGED FALSE POSITIVE RESULTS. CUSTOMER RETESTED USING A DIFFERENT PLATFORM (ALINITY & GENEXPERT). THE LOT NUMBER OF THE ASSAY USED WAS NOT PROVIDED TO HOLOGIC.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
390973 FUSION SARS-COV-2 REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID OJR HOLOGIC, INC.

Patients

Seq Age Sex Outcome Treatment
1 Unknown