FDA Adverse Event Malfunction Summary report: N

SPECTRANETICS 14F CAVACLEAR LASER SHEATH

MDR report key: 15583738 · Received October 11, 2022

Report

Report Number
1721279-2022-00175
Event Type
Malfunction
Date Received
October 11, 2022
Date of Event
September 16, 2022
Report Date
October 11, 2022
Manufacturer
THE SPECTRANETICS CORPORATION
Product Code
QRJ
UDI-DI
00813132022969
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PMA/510K: 510K/PMA NUMBER IS NOT APPLICABLE. DE NOVO NUMBER: (B)(4). DEVICE EVALUATED BY MFR: THE DEVICE WAS DISCARDED, THUS NO INVESTIGATION COULD BE COMPLETED. BASED ON THE COMPLAINT DETAILS, EXCESSIVE FORCE WAS APPLIED DURING USE. SUBMISSION OF THIS REPORT DOES NOT, IN ITSELF, REPRESENT A CONCLUSION BY THE MANUFACTURER AND/OR AUTHORIZED REPRESENTATIVE OR THE NATIONAL COMPETENT AUTHORITY THAT THE CONTENT OF THIS REPORT IS COMPLETE OR ACCURATE, THAT THE MEDICAL DEVICE(S) LISTED FAILED IN ANY MANNER AND/OR THAT THE MEDICAL DEVICE(S) CAUSED OR CONTRIBUTED TO AN ALLEGED DEATH OR DETERIORATION IN THE STATE OF THE HEALTH OF ANY PERSON.

Description of Event or Problem · 0

A PROCEDURE COMMENCED TO REMOVE AN INFERIOR VENA CAVA (IVC) FILTER. THE FILTER WAS NO LONGER NEEDED, AND REMOVING IT WOULD ELIMINATE THE NEED FOR ANTICOAGULATION. THE PHYSICIAN CHOSE A SPECTRANETICS 14F CAVACLEAR LASER SHEATH TO REMOVE THE FILTER. AS THE CAVACLEAR WAS BEING USED, EXCESSIVE PUSH/PULL FORCE WAS APPLIED BY THE PHYSICIAN. AS A RESULT, THE AREA WHERE THE DEVICE''S BIFURCATE HANDLE AND THE WORKING LENGTH MEET, BENT AT A 90 DEGREE ANGLE, EXPOSING THE FIBERS. THE PHYSICIAN DISCONTINUED USE OF THE FIRST CAVACLEAR, OBTAINED A NEW CAVACLEAR, AND THE FILTER WAS SUCCESSFULLY REMOVED WITH NO REPORTED PATIENT HARM. THIS EVENT IS BEING REPORTED FOR UNINTENDED RADIATION EXPOSURE, POTENTIAL FOR HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2943091 SPECTRANETICS 14F CAVACLEAR LASER SHEATH LASER-POWERED INFERIOR VENA CAVA FILTER RETRIEVAL QRJ THE SPECTRANETICS CORPORATION 500-514 C6P21J23A 00813132022969

Patients

Seq Age Sex Outcome Treatment
1 42 YR Female COOK MEDICAL GUNTHER TULIP VENA CAVA FILTER| SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM