BD FACS¿ LYSE WASH ASSISTANT
Report
- Report Number
- 2916837-2022-00294
- Event Type
- Malfunction
- Date Received
- October 11, 2022
- Date of Event
- September 29, 2022
- Report Date
- October 5, 2022
- Manufacturer
- BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
- Product Code
- JQW
- UDI-DI
- 00382903371464
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT WHILE USING THE BD FACS¿ LYSE WASH ASSISTANT THAT THE CUSTOMER OBSERVED CONTAMINATION OF TUBES IN FOLLOWING TUBES BEING PROCESSED BY THE LWA. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER OBSERVED CONTAMINATION OF TUBES IN FOLLOWING TUBES BEING PROCESSED BY THE LWA. HE PERFORMED SEVERAL TESTS IN ORDER TO CALCULATE THE CARRYOVER PERCENTAGE, AND THIS CARRYOVER SEEMS TO BE HIGHER THAN THE MAXIMUM CARRYOVER THAT IS SPECIFIED IN THE TECHNICAL SPECIFICATIONS OF THE INSTRUMENT. AS THERE ANY ALLEGED INJURY OR HARM TO USER OR PATIENT? NO. SERIOUS INJURY, NO. ERRONEOUS RESULTS, POTENTIALLY YES. COURSE OF TREATMENT CHANGED, NO. EXPOSURE TO BLOOD/BODILY FLUID NO. SAFETY ISSUE INCL. NEEDLE STICK, MEDICAL INTERVENTION, NO.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2483292 | BD FACS¿ LYSE WASH ASSISTANT | STATION, PIPETTING DILUTING CLINICAL USE | JQW | BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES | 337146 | 00382903371464 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |