FDA Adverse Event Death Summary report: N

TCI IP LEFT VENTRICULAR ASSIST SYSTEM

MDR report key: 155691 · Received March 11, 1998

Report

Report Number
1218444-1998-00005
Event Type
Death
Date Received
March 11, 1998
Date of Event
February 7, 1998
Report Date
February 10, 1998
Manufacturer
THERMO CARDIOSYSTEMS INC.
Product Code
DSQ
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

021098-PHYSICIAN THOUGHT THERE MIGHT BE A TECHNICAL PROBLEM SINCE PATIENT EXPERIENCED ACUTE FLOW DROP. PATIENT TAKEN TO O.R.; AIR WAS "EVERYWHERE" AND "IN THE AORTA" DECISION MADE TO TERMINATE EVERYTHING. 021198-PHYSICIAN STATED THAT PATIENT HAD A BSA OF 1.8 AND A LARGE HEART. THE DEVICE WAS IN THE NORMAL POSITION. THE INLET MAY BE TOO HIGH, THE UMBILICUS WAS HIGH. PERC LINE PUT RIGHT OF UMBILICUS. SEWING RING ATTACHED TO INFLOW CONDUIT WITH ATTACHED BLUE THREAD, THE BAND AND UMBILICAL TAPE. PATIENT TAMPONADE. WHEN OPENED IN O.R., AIR SUCKED. INLET CONDUIT FOUND TO BE OUT OF SEWING RING. DEVICE WAS NOT ANCHORED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TCI IP LEFT VENTRICULAR ASSIST SYSTEM Implant LEFT VENTRICULAR ASSIST DEVICE (LVAD) DSQ THERMO CARDIOSYSTEMS INC. 1000A NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Death