ENDOPATH ETS
Report
- Report Number
- 1527736-1998-00681
- Event Type
- Malfunction
- Date Received
- March 11, 1998
- Report Date
- February 10, 1998
- Manufacturer
- ETHICON ENDO-SURGERY, INC.
- Product Code
- KOG
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER
Narratives
THE PRODUCT COMPLAINT ANALYSIS TEAM HAS COMPLETED ITS INVESTIGATION OF THE DEVICE. VISUAL INSPECTIONS & RESULTS: BATCH NUMBER, K01067; CARTIDGE PAN IN PLACE/CONDITION, YES/GOOD; CONDITION OF DRIVERS, GOOD; LOCKOUT TABS ON PAN CONDITION, BENT; POSITION/CONDITION OF WEDGE SLEDS, PARTIALLY FIRED. FUNCTIONAL TESTS & RESULTS: CONDITION OF CLAMP FIRST LOCKOUT, GOOD; CONDITION OF CLAMPING MECHANISM, GOOD; CONDITION OF FIRING MECHANISM, GOOD; CONDITION OF KNIFE, GOOD; CONDITION OF WEDGE BANDS, GOOD; IS HYPER LOCKOUT CONDITION PRESENT, NO; RESULT OF ATTEMPTED FIRING, GOOD. ANALYSIS CONCLUSION: BASED UPON THE INQUIRY INFO REC'D, THE VISUAL EXAMINATION, AND THE FUNCTIONAL TESTING, NO CONCLUSION COULD BE REACHED AS TO WHY THE INSTRUMENT REPORTEDLY "WOULD NOT CUT" DURING SURGERY. THE INSTRUMENT WAS RETURNED IN GOOD PHYSICAL CONDITION. THE INSTRUMENT WAS CYCLED, FIRED, CUT, AND FORMED THE STAPLES WITHIN DESIGN SPEC. IT WAS CONCLUDED THAT THE INSTRUMENT WAS FULLY FUNCTIONAL AND CONFORMING TO DESIGN SPECS. THE EXPERIENCE THE SURGEON REPORTED COULD NOT BE REPEATED. THE RETURNED CARTRIDGE HAD A BENT LOCKOUT TAB ON THE PAN WHICH INDICATES THAT THE INSTRUMENT'S FIRING CYCLE WAS INTERRUPTED, RELEASED, THEN RESTARTED. WHEN THIS OCCURRED, THE LOCKOUT TAB ON THE CARTRIDGE BECAME DAMAGED. IF THE INSTRUMENT'S FIRING CYCLE IS INTERRUPTED, RELEASED, THEN RESTARTED, THE CARTRIDGE WILL LOCKOUT AND A NEW CARTRIDGE SHOULD BE LOADED INTO THE INSTRUMENT. EACH INSTRUMENT IS EVALUATED DURING THE ASSEMBLY PROCESS TO ENSURE IT FUNCTIONS PROPERLY.
IT WAS REPORTED DURING A LAPAROSCOPIC ABDOMINAL EXPLORATION WITH INDENTAL LAP APPENDECTOMY THE TSB35 WOULD NOT CUT TISSUE, BUT STAPLED FINE. ALTHOUGH TSW35 WAS OPENED TO FINISH THE CASE. THERE WAS NO CONSEQUENCE TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH ETS | LINEAR CUTTER | KOG | ETHICON ENDO-SURGERY, INC. | NA | K47955 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |