FDA Adverse Event Injury Summary report: N

TECNIS IOL

MDR report key: 15558706 · Received October 7, 2022

Report

Report Number
3012236936-2022-02541
Event Type
Injury
Date Received
October 7, 2022
Date of Event
December 7, 2021
Report Date
October 31, 2022
Manufacturer
AMO MANUFACTURING NETHERLANDS
Product Code
MFK
UDI-DI
05050474750685
PMA / PMN Number
P980040
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS NOT RETURNED FOR EVALUATION. THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. A REVIEW OF THE DEVICE HISTORY RECORD, COMPLAINT TRENDING, AND RISK DOCUMENTATION FOR THIS DEVICE WILL BE PERFORMED. UPON COMPLETION OF THE REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION A SUPPLEMENTAL MEDWATCH WILL BE FILED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

SECTION D9: DEVICE AVAILABLE FOR EVALUATION: YES, SECTION D9: RETURNED TO MANUFACTURER ON: OCT 13, 2022, SECTION H3: DEVICE EVALUATED BY MANUFACTURER: YES. DEVICE EVALUATION: THE PRODUCT WAS RETURNED TO THE MANUFACTURING SITE FOR EVALUATION. VISUAL INSPECTION UNDER MAGNIFICATION REVEALED THAT THE COMPLAINT LENS WAS RECEIVED CUT IN HALF AND COATED IN VISCOELASTIC RESIDUE. THE LENS WAS CLEANED AND, NO ISSUES THAT COULD CONTRIBUTE TO OR CAUSE BLURRY VISION WERE IDENTIFIED. BASED ON THE RETURN CONDITION OF THE LENS, NO FURTHER PRODUCT EVALUATION CAN BE PERFORMED. CONCLUSION: THE REPORTED COMPLAINT ISSUE WAS NOT CONFIRMED. THE OTHER OBSERVED ISSUES DURING THE PRODUCT EVALUATION COULD NOT BE CONFIRMED TO BE RELATED TO A MANUFACTURING OR DESIGN ISSUE. THEREFORE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY OR PRODUCT MALFUNCTION. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PRELOADED INTRAOCULAR LENS (IOL) WAS EXPLANTED FROM THE PATIENT'S RIGHT EYE DUE TO BLURRY VISION. THE NEAR VISION OF THE PATIENT WAS BLURRIER. THE IOL WAS EXPLANTED AND ANOTHER COMPETITOR LENS WAS USED AS THE REPLACEMENT LENS. THE PATIENT WAS DOING GOOD POST-OPERATION. NO OTHER INTERVENTION WAS REQUIRED. NO OTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2779168 TECNIS IOL LENS, MULTIFOCAL INTRAOCULAR MFK AMO MANUFACTURING NETHERLANDS DFR00V 05050474750685

Patients

Seq Age Sex Outcome Treatment
1 55 YR Female Required Intervention