FDA Adverse Event Injury Summary report: N

AK 96 MACHINES

MDR report key: 15549983 · Received October 6, 2022

Report

Report Number
9616026-2022-00099
Event Type
Injury
Date Received
October 6, 2022
Date of Event
September 14, 2022
Report Date
November 9, 2022
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
KDI
PMA / PMN Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(6). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THE ACTUAL DEVICE WAS EVALUATED ON SITE BY A LOCAL HOSPITAL ENGINEER. DURING INSPECTION A LOOSENED TUBING THAT CONNECT THE MACHINE TO THE DIALYZER FILTER THAT CAUSED THE EXTERNAL LEAK WAS FOUND. THIS WAS REPAIRED AND AFTER RECONNECTING THE LOOSEN CONNECTOR, THE MACHINE WAS BACK TO NORMAL. THE CAUSE OF THE CONDITION COULD NOT BE DETERMINED. A SERVICE HISTORY REVIEW REVEALED NO INDICATION THAT THE PARTS REPLACED DURING SERVICING CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT APPROXIMATELY ONE (1) HOUR AFTER STARTING HEMODIALYSIS (HD) THERAPY USING AN AK 96 CONTROL UNIT, THE PATIENT EXPERIENCED FAINTING, VOMITING AND SHOCK. THE PATIENT WEIGHT WAS 51.1 KG G AFTER (ONE) 1 HOUR OF TREATMENT INDICATING AN EXCESSIVE FLUID REMOVAL OF 4.8L FROM THE START WEIGHT. THE PATIENT WAS TREATED WITH OXYGEN ADMINISTRATION, 2000 ML NORMAL SALINE, 5 MG DEXAMETHASONE, 20 MG OF GLUCOSE. THE PATIENT RECOVERED FROM THE EVENT. NO ALARM WAS GENERATED DURING THE TREATMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2889220 AK 96 MACHINES DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM KDI BAXTER HEALTHCARE CORPORATION NA NA

Patients

Seq Age Sex Outcome Treatment
1 53 YR Male Required Intervention POLYFLUX 14L DIALYZER| SANXIN A- CONCENTRATE| SANXIN BLOOD LINES