FDA Adverse Event Malfunction Summary report: N

THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER

MDR report key: 15544120 · Received October 5, 2022

Report

Report Number
2029046-2022-02404
Event Type
Malfunction
Date Received
October 5, 2022
Date of Event
July 14, 2022
Report Date
October 5, 2022
Manufacturer
BIOSENSE WEBSTER INC
Product Code
LPB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE BIOSENSE WEBSTER, INC. PRODUCT ANALYSIS LAB RECEIVED THE DEVICE ON 29-AUG-2022. THE DEVICE EVALUATION WAS COMPLETED ON 06-SEP-2022. THE THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER WAS RETURNED TO BIOSENSE WEBSTER (BWI) FOR EVALUATION. A VISUAL INSPECTION, TEMPERATURE AND IMPEDANCE TEST OF THE RETURNED DEVICE WERE PERFORMED FOLLOWING WITH BWI PROCEDURES. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THE PEBAX WAS BROKEN. THE TEMPERATURE AND IMPEDANCE TEST WERE PERFORMED, AND NO IMPEDANCE WAS DISPLAYED DUE TO AN OPEN CIRCUIT ON THE TIP AREA. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE 30672380M NUMBER, AND NO INTERNAL ACTION WAS FOUND DURING THE REVIEW. THE ISSUE REPORTED BY THE CUSTOMER WAS CONFIRMED. IT SHOULD BE NOTED THAT PRODUCT FAILURE IS MULTIFACTORIAL. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING WARNING STATED IN THE CARTO 3 SYSTEM MANUAL: IF THE RF GENERATOR DOES NOT DISPLAY TEMPERATURE, VERIFY THAT THE APPROPRIATE CABLE IS PLUGGED INTO THE RF GENERATOR. AS PART OF BIOSENSE WEBSTER'S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THIS PRODUCT ISSUE WILL BE ADDRESSED THROUGH BWI'S QUALITY SYSTEM. THE ¿SUSPECTED MEDICAL DEVICE¿ REPORTED IN SECTION D OF THIS REPORT IS NOT MARKETED IN USA OR APPROVED BY THE FDA. HOWEVER, IT IS BEING REPORTED AS BIOSENSE WEBSTER CONSIDERS THIS AS A SIMILAR DEVICE TO THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER(D133601) APPROVED UNDER PMA # P030031/S053. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 4. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY BIOSENSE WEBSTER, INC., OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, BIOSENSE WEBSTER, INC., OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. MANUFACTURER'S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL TACHYCARDIA (AT) PROCEDURE WITH A THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER AND THE BIOSENSE WEBSTER INC, (BWI) PRODUCT ANALYSIS LAB OBSERVED A BROKEN PEBAX. INITIALLY, IT WAS REPORTED THAT DURING THE OPERATION, THERE WAS NO TEMPERATURE DISPLAYED ON THE CARTO OR GENERATOR. A SECOND CATHETER WAS USED TO COMPLETE THE OPERATION. THERE WAS NO ADVERSE EVENT REPORTED ON PATIENT. THE ISSUE OF NO TEMPERATURE WAS ASSESSED AS NOT MDR REPORTABLE. THE POTENTIAL THAT IT COULD CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY, OR OTHER SIGNIFICANT ADVERSE EVENT, IS REMOTE. THIS EVENT IS BEING REPORTED BECAUSE THE BIOSENSE WEBSTER INC, (BWI) PRODUCT ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION AND VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THE PEBAX WAS BROKEN. THIS FINDING WAS ASSESSED AS MDR REPORTABLE. THEREFORE, THE AWARENESS DATE FOR THIS REPORTABLE LAB FINDING IS 06-SEP-2022.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2080554 THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER CARDIAC ABLATION PERCUTANEOUS CATHETER LPB BIOSENSE WEBSTER INC 30672380M

Patients

Seq Age Sex Outcome Treatment
1 Unknown UNK BRAND CATHETER.| UNK_CARTO 3.