FDA Adverse Event Injury Summary report: N

TALENT TAA STENT GRAFT

MDR report key: 15543637 · Received October 5, 2022

Report

Report Number
2953200-2022-00022
Event Type
Injury
Date Received
October 5, 2022
Date of Event
January 26, 2009
Report Date
October 5, 2022
Manufacturer
MEDTRONIC CARDIOVASCULAR SANTA ROSA
Product Code
MIH
PMA / PMN Number
P070007
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC RECEIVED THE FOLLOWING INFORMATION FROM A JOURNAL ARTICLE ENTITLED; RETROGRADE TYPE A AORTIC DISSECTION AFTER ENDOVASCULAR STENT GRAFT PLACEMENT FOR TREATMENT OF TYPE B DISSECTION. DONG ZH, FU WG, WANG YQ, GUO DQ, XU X, JI Y, CHEN B, JIANG JH, YANG J, SHI ZY, ZHU T, SHI Y. CIRCULATION. 2009;119 (5) PP. 735-741 DOI: 10.1161/CIRCULATIONAHA.107.759076. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

A TALENT STENT GRAFT WAS IMPLANTED IN A PATIENT IN THE ENDOVASCULAR TREATMENT OF A TYPE B AORTIC DISSECTION ON AN UNKNOWN DATE OVER A 7 YEAR PERIOD. THE STENT GRAFT WAS PLACED IN THE DISTAL AORTIC ARCH WITH > 20-MM PROXIMAL LANDING ZONE AND 10 % STENT GRAFT OVERSIZING. THE INITIAL INDICATION FOR THE TREATMENT OF THE TYPE B AORTIC DISSECTION WAS REFRACTORY HYPERTENSION. IT WAS REPORTED THAT INTRA-OPERATIVELY, A TYPE A AORTIC DISSECTION WAS DIAGNOSED AT THE LOCATION OF THE PROXIMAL BARE SPRING¿COVERING AREA. THE RTAD WAS ASCRIBED TO A CUTTING INJURY BY THE RADIOPAQUE GOLD MARKERS OF THE ANGIOGRAPHIC CATHETER PLACED IN THE ASCENDING THORACIC AORTA THROUGH THE LEFT SUBCLAVIAN ARTERY ACROSS WHICH THE PROXIMAL BARE SPRING WAS INTENTIONALLY PLACED THAT THUS PRESSED THE CATHETER AGAINST THE AORTIC WALL. CONSEQUENTLY, WHEN THE CATHETER WAS PULLED BACK, THE GOLD MARKER PROTRUDING A LITTLE OVER THE CATHETER SURFACE GENERATED A CUTTING FORCE ON THE GREATER CURVE, AND A NEW TEAR FORMED. THIS RTAD WAS TREATED MY MEDICAL MANAGEMENT. THE PATIENT WAS THEN LOST TO FOLLOW-UP 16 DAYS LATER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2126629 TALENT TAA STENT GRAFT SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR MIH MEDTRONIC CARDIOVASCULAR SANTA ROSA UNK-CV-SR-TAL TAA

Patients

Seq Age Sex Outcome Treatment
1 35 YR Unknown Required Intervention| O