SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2009-03091
- Event Type
- Injury
- Date Received
- December 9, 2009
- Date of Event
- September 14, 2009
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: THE ACTUAL LEAD WAS NOT RECEIVED FOR EVALUATION. PERFORMANCE DATA COLLECTED FROM THE ICD DEVICE INDICATES THE RV PACE IMPEDANCE MEASUREMENT HAS BEEN VARIABLE THROUGHOUT THE RECORD ENDING 25 AUG 2009. THE MIN VALUE RANGED FROM 424 - 1200 OHMS AND THE MAX VALUE RANGED FROM 560 - 1568 OHMS. THE LIFETIME VENTRICULAR SENSING INTEGRITY COUNTER IS 1193 AND ALL COUNTS OCCURRED SINCE 28 MAY 2008. A HIGH VALUE OF THIS COUNT CAN INDICATE A POSSIBLE INTERMITTENCY IN THE SYSTEM.
IT WAS REPORTED THAT THE LEAD HAD VARYING IMPEDANCE, HIGH IMPEDANCE, NOISE/OVERSENSING, AND POSSIBLE FRACTURE. ALSO REPORTED WAS THAT THE LEAD WAS "VISIBLY POKING FROM WITHIN THE POCKET." A NEW RV LEAD WAS IMPLANTED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Required Intervention | D154VWC IMPLANTABLE PACEMAKER/CARDIO/DEFIB |