FDA Adverse Event Malfunction Summary report: N

ELECTROSURGICAL GENERATOR "ESG-400"

MDR report key: 15541848 · Received October 5, 2022

Report

Report Number
9610773-2022-00458
Event Type
Malfunction
Date Received
October 5, 2022
Date of Event
September 22, 2022
Report Date
October 5, 2022
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
GEI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE SUSPECT MEDICAL DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR INVESTIGATION BUT TO OLYMPUS SHANGHAI (OSH) IN CHINA (RETURNED TO OSH ON (B)(6) 2022). THE EVALUATION CONFIRMED THE OCCURRENCE OF ERROR MESSAGE E433 WHICH IS TRIGGERED BY THE GENERATOR¿S SAFETY SYSTEM. IN CASE OF CRITICAL ERRORS, THE SAFETY SYSTEM WILL NOT PERMIT ANY FURTHER USE OF THE GENERATOR UNTIL THE ERROR IS RECTIFIED. IN THE CASE AT HAND, THIS WAS CAUSED BY A DEFECTIVE GENERATOR BOARD. THUS, THIS EVENT/INCIDENT WAS ATTRIBUTED TO COMPONENT FAILURE. A MANUFACTURING AND QUALITY CONTROL REVIEW WAS PERFORMED FOR THE AFFECTED SERIAL NUMBER OF THE HF GENERATOR WITHOUT SHOWING ANY ABNORMALITIES. THE CASE WILL BE CLOSED ON OLYMPUS SIDE WITH NO FURTHER ACTIONS, BUT THE REPORTED EVENT/INCIDENT WILL BE RECORDED FOR TRENDING AND SURVEILLANCE PURPOSES. FURTHERMORE, THE USER WILL BE INFORMED ABOUT THE INVESTIGATION RESULTS. PLEASE NOTE: THIS REPORT IS BEING SUBMITTED ALTHOUGH THE SUSPECT MEDICAL DEVICE IS NOT MARKETED IN THE USA. HOWEVER, A SIMILAR DEVICE IS MARKETED. MODEL # / CATALOG #: WB91051W; BRAND NAME: ELECTROSURGICAL GENERATOR "ESG-400"; COMMON DEVICE NAME: HF-GENERATORS; 510(K): K203682; PRODUCT CODE: GEI.

Description of Event or Problem · 0

OLYMPUS WAS INFORMED THAT DURING REGULAR MAINTENANCE AT THE USER FACILITY, THE ELECTROSURGICAL GENERATOR ESG-400 DISPLAYED ERROR MESSAGE E433 AND AUTO-STARTED ITSELF CONTINUOUSLY. NO PROCEDURE NOR PATIENT WAS INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1560257 ELECTROSURGICAL GENERATOR "ESG-400" HF-GENERATORS GEI OLYMPUS WINTER & IBE GMBH WB91051C

Patients

Seq Age Sex Outcome Treatment
1 Unknown