FDA Adverse Event
Malfunction
Summary report: N
FOCUS RADIATION TREATMENT PLANNING SYSTEM
MDR report key: 155400
·
Received March 9, 1998
Report
- Report Number
- 1937649-1998-00002
- Event Type
- Malfunction
- Date Received
- March 9, 1998
- Date of Event
- February 23, 1998
- Report Date
- March 5, 1998
- Manufacturer
- COMPUTERIZED MEDICAL SYSTEMS, INC.
- Product Code
- KPQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
USER REPORTED THAT WHEN VALIDATING THE PHYSICS DATA, AS REQUIRED BY THE FOCUS RTP SYSTEM, THEY NOTED THE RELATIVE DENSITY FIGURES IN THE PHYSICS DATA TABLE FOR SELECTED MATERIALS WERE INCORRECT. THE PROBLEM WAS CREATED BY CMS LOADING INCORRECT NUMBERS INTO THE TABLE. THERE IS A WORK-AROUND FOR CUSTOMERS. NO PTS WERE TREATED. THIS ONLY AFFACTS PHOTON DOSE CALCULATION DONE USING THE FFT ALGORITHM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FOCUS RADIATION TREATMENT PLANNING SYSTEM | RTP: SYSTEM | KPQ | COMPUTERIZED MEDICAL SYSTEMS, INC. | 9200 | RELEASE 2.0.0 ONLY |