FDA Adverse Event Malfunction Summary report: N

SPECTRUM IQ INFUSION PUMP

MDR report key: 15533842 · Received October 3, 2022

Report

Report Number
1314492-2022-04247
Event Type
Malfunction
Date Received
October 3, 2022
Report Date
November 11, 2022
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
UDI-DI
00085412610900
PMA / PMN Number
K220417
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THE DEVICE WAS RECEIVED FOR EVALUATION. FUNCTIONAL TESTING WAS PERFORMED AND DID NOT IDENTIFY ANY ISSUES RELATED TO THE CUSTOMER REPORTED CONDITION. FLOW RATE TESTING WAS PERFORMED, AND THE FLOW RATE OF THE DEVICE WAS FOUND TO BE WITHIN SPECIFICATION. A SERVICE HISTORY REVIEW WAS PERFORMED AND REVEALED THAT THE DEVICE HAS NO PREVIOUS SERVICE EVENTS; THEREFORE, SERVICING DID NOT CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE CONDITION WAS NOT DETERMINED. THE CONDITION MAY BE RELATED TO FA -2021-056, WHICH IS ASSOCIATED WITH REINFORCING PROPER INTRAVENOUS LINE SETUP AND DETECTION OF UPSTREAM OCCLUSIONS FOR SPECTRUM PUMPS. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT A SPECTRUM IQ PUMP UNDER INFUSED DURING THERAPY AT THE INTENSIVE CARE UNIT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.