SPECTRUM IQ INFUSION PUMP
Report
- Report Number
- 1314492-2022-04247
- Event Type
- Malfunction
- Date Received
- October 3, 2022
- Report Date
- November 11, 2022
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- UDI-DI
- 00085412610900
- PMA / PMN Number
- K220417
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
THE DEVICE WAS RECEIVED FOR EVALUATION. FUNCTIONAL TESTING WAS PERFORMED AND DID NOT IDENTIFY ANY ISSUES RELATED TO THE CUSTOMER REPORTED CONDITION. FLOW RATE TESTING WAS PERFORMED, AND THE FLOW RATE OF THE DEVICE WAS FOUND TO BE WITHIN SPECIFICATION. A SERVICE HISTORY REVIEW WAS PERFORMED AND REVEALED THAT THE DEVICE HAS NO PREVIOUS SERVICE EVENTS; THEREFORE, SERVICING DID NOT CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE CONDITION WAS NOT DETERMINED. THE CONDITION MAY BE RELATED TO FA -2021-056, WHICH IS ASSOCIATED WITH REINFORCING PROPER INTRAVENOUS LINE SETUP AND DETECTION OF UPSTREAM OCCLUSIONS FOR SPECTRUM PUMPS. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A SPECTRUM IQ PUMP UNDER INFUSED DURING THERAPY AT THE INTENSIVE CARE UNIT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1700717 | SPECTRUM IQ INFUSION PUMP | PUMP, INFUSION | FRN | BAXTER HEALTHCARE CORPORATION | NA | NA | 00085412610900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |