BD ULTRA-FINE¿ INSULIN SYRINGES
Report
- Report Number
- 1920898-2022-00673
- Event Type
- Malfunction
- Date Received
- October 3, 2022
- Date of Event
- September 12, 2022
- Report Date
- October 27, 2022
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DATE OF EVENT IS UNKNOWN; AWARENESS DATE HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY: NO SAMPLES WERE RETURNED THEREFORE THE INVESTIGATION WAS PERFORMED BASED ON THE PHOTOS PROVIDED. THREE PHOTOS OF 0.3ML BD INSULIN SYRINGES FROM LOT# 0286306 WERE PROVIDED. THE CUSTOMER REPORTED THAT THERE IS LIQUID IN THE SYRINGES. THE PHOTOS WERE EXAMINED, AND IT WAS OBSERVED THAT A TRANSPARENT MATERIAL WAS PRESENT INSIDE THE SYRINGE BARREL. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 0020548. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. EMBECTA WAS ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE BASED ON THE PHOTOS RECEIVED (FM). A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED.
IT WAS REPORTED WHILE USING BD ULTRA-FINE¿ INSULIN SYRINGES FOREIGN MATTER WAS FOUND IN THE FLUID PATHWAY. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MY INSULIN SYRINGES SOMETIMES HAVE A CLEAR LIQUID EQUIVALENT TO A DROP OF ONE MILLIMETER IN DIAMETER.
IT WAS REPORTED WHILE USING BD ULTRA-FINE¿ INSULIN SYRINGES FOREIGN MATTER WAS FOUND IN THE FLUID PATHWAY. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MY INSULIN SYRINGES SOMETIMES HAVE A CLEAR LIQUID EQUIVALENT TO A DROP OF ONE MILLIMETER IN DIAMETER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1768475 | BD ULTRA-FINE¿ INSULIN SYRINGES | PISTON SYRINGE | FMF | BD MEDICAL - DIABETES CARE | 0020548 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |