FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® LH PST¿ II PLUS BLOOD COLLECTION TUBES

MDR report key: 15525463 · Received October 1, 2022

Report

Report Number
9617032-2022-00966
Event Type
Malfunction
Date Received
October 1, 2022
Date of Event
September 23, 2022
Report Date
September 23, 2022
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: "MATERIAL #: 367376. LOT/BATCH #: 2152349. BD HAD NOT RECEIVED SAMPLES, BUT 1 PHOTO WAS PROVIDED FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND THE INDICATED FAILURE MODE FOR FOREIGN MATTER - INSECT WAS OBSERVED. BD WAS NOT ABLE TO DETERMINE WHERE OR WHEN THE INSECT MAY HAVE ENTERED THE TUBE. THE MANUFACTURING SITE UTILIZES PEST BAIT BOXES, ELECTRONIC FLYING INSECT KILLERS, AND IS VISITED BY A FIELD BIOLOGIST TWICE A YEAR TO REVIEW OUR SYSTEMS AND HELP PREVENT PESTS. ADDITIONALLY, (B)(4) RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND THE ISSUE OF FOREIGN MATTER - INSECT WAS NOT OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE FOREIGN MATTER - INSECT. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS."

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD VACUTAINER® LH PST¿ II PLUS BLOOD COLLECTION TUBES CUSTOMER FOUND A FLY IN THE TUBE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: ¿FINDING A FLY IN THE TUBE.¿

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2337056 BD VACUTAINER® LH PST¿ II PLUS BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 2152349

Patients

Seq Age Sex Outcome Treatment
1 Unknown