FDA Adverse Event
Malfunction
Summary report: N
T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
MDR report key: 15519588
·
Received September 30, 2022
Report
- Report Number
- 3013756811-2022-108006
- Event Type
- Malfunction
- Date Received
- September 30, 2022
- Date of Event
- September 8, 2022
- Report Date
- November 14, 2022
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00850006613762
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DC, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
Additional Manufacturer Narrative · 0
THE DEVICE HAS BEEN RECEIVED FOR EVALUATION; HOWEVER, DEVICE EVALUATION IS NOT YET COMPLETE. A SUPPLEMENTAL REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION.
Description of Event or Problem · 0
IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 160-208 MG/DL. REPORTEDLY, THE CUSTOMER DID NOT HAVE A METHOD OF BACKUP INSULIN THERAPY AND DECLINED FOLLOW-UP TO ENSURE BACKUP THERAPY WAS OBTAINED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2171365 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1002717 | 00850006613762 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR | Female |