FDA Adverse Event Injury Summary report: N

KYPHX HV-R BONE CEMENT

MDR report key: 1550294 · Received December 2, 2009

Report

Report Number
2953769-2009-00184
Event Type
Injury
Date Received
December 2, 2009
Date of Event
January 1, 2009
Report Date
November 9, 2009
Manufacturer
MEDTRONIC SPINE LLC.
Product Code
NDN
PMA / PMN Number
K033801
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

REPORT SOURCE. EUR SPINE J (2009) 18:1713-1826, PAGE 1771 - P 72 - ABSTRACT TITLED "INTRADURAL PMMA LEAKAGE AND INCOMPLETE CONUS SYNDROME AFTER PERCUTANEOUS KYPHOPLASTY WITHOUT DURAL LESION: CASE REPORT" BY G. BRINKER, M. REINER, W. VON TEMPELHOFF, R.-I. ERNESTUS, R. GOLDBRUNNER. DEVICE NOT RETURNED, FOLLOW UP WITH AUTHOR.

Description of Event or Problem · 1

IN AN ABSTRACT TITLED "INTRADURAL PMMA LEAKAGE AND INCOMPLETE CONUS SYNDROME AFTER PERCUTANEOUS KYPHOPLASTY WITHOUT DURAL LESION: CASE REPORT", THE FOLLOWING EVENT WAS REPORTED: "A FEMALE SUFFERING FROM A FRACTURE OF THE 12TH THORACAL VERTEBRA WAS TREATED BY PERCUTANEOUS TRANSPEDICULAR KYPHOPLASTY WITH PMMA. POSTOPERATIVELY, AN EPIDURAL HEMATOMA WAS EVACUATED. DURING THE FOLLOWING MONTHS, THE PATIENT PRESENTED A PROGREDIENT CONUS/CAUDA EQUINA SYNDROME. MYELOGRAPHIC CT AND MRI SHOWED AN INTRADURAL SPACE OCCUPYING LESION NEXT TO THE CONUS, WHICH WAS ACCORDING TO ITS SIGNAL INTENSITY SUSPECTED TO BE PMMA. SURGERY WAS PERFORMED BY HEMILAMINECTOMY OF T-12 AND INTRADURAL DECOMPRESSION OF THE CONUS/CAUDA. AN INTRADURAL PMMA-MASS (15X5X4 MM) NEXT TO THE CONUS WAS REMOVED. BASED ON EVIDENCE, IT IS ASSUMED THAT THE PMMA LEAKAGE WAS DUE TO TRANSVENOUS INFLOW VIA EPIDURAL-INTRADURAL ANASTOMOSIS. THREE MONTHS POSTOPERATIVELY THE CLINICAL STATE IMPROVED MARKEDLY, BUT INCOMPLETELY." NO FURTHER INFORMATION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 KYPHX HV-R BONE CEMENT HIGH VISCOSITY, RADIOPAQUE BONE CEMENT NDN MEDTRONIC SPINE LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 58 YR Other| R