X-SMART IQ BASIC STARTER KIT
Report
- Report Number
- 8031010-2022-00645
- Event Type
- Malfunction
- Date Received
- September 28, 2022
- Report Date
- September 29, 2022
- Manufacturer
- MAILLEFER INSTRUMENTS HOLDING SARL
- Product Code
- EBW
- PMA / PMN Number
- K161213
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
ADDITIONAL INFORMATION HAS BEEN RECEIVED THAT THERE WAS NO INJURY THAT OCCURRED FOR THIS EVENT. THIS FOLLOW UP REPORT IS FOR THIS ADDITIONAL INFORMATION.
WHILE THERE IS NO INDICATION THAT SERIOUS INJURY RESULTED IN THIS EVENT, THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. INVESTIGATION SUMMARY: RECEIVED = 1X X-SMART IQ CONTRA-ANGLE A105100000000 SN: (B)(4). 1X X-SMART IQ HANDPIECE H105900000000 SN: (B)(4). 1X X-SMART IQ UNIVERSAL CHARGER A105300000000 1X PROPEX IQ X-SMART IQ MEASUREMENT CABLE B00PPIQ1ACHPA 1X PROPEX IQ PROPEX IQ CHARGER B00PPIQ1ACCHG 1X PROPEX IQ PROPEX IQ FILE CLIP X2 B00PPIQ2ACFCL 1X PROPEX IQ PROPEX IQ UNIT BSPPPIQ000002 SN: (B)(4). X-SMART IQ CONTRA-ANGLE. SAV VARIOUS MECHANICAL PROBLEM. NOISY, NOT FLUID ROTATION. X-SMART IQ HANDPIECE. SAV OTHER. CONNECTOR BROKEN. PROPEX IQ PROPEX IQ UNIT. SAV DEFECTIVE BATTERY. BATTERY DISCHARGING TOO FAST. PROPEX IQ X-SMART IQ MEASUREMENT CABLE. SAV VARIOUS CABLE PROBLEM. BAD CONTACT. ALL COMPLAINTS ARE MONITORED THROUGH THE MONTHLY PRODUCT SURVEILLANCE COMMITTEE AND A CORRECTIVE ACTION COULD BE DETERMINED BY THE COMMITTEE.
IN THIS EVENT IT IS REPORTED THAT X-SMART IQ BASIC STARTER KIT WAS NOT HAVING A FLUID ROTATION AND WHILE IN USE WITH PROPEX IQ APEX LOCATOR THE DEVICE WAS GIVING INCORRECT MEASUREMENTS. OUTCOME OF THIS EVENT IS UNKNOWN AS OF THIS MDR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2302219 | X-SMART IQ BASIC STARTER KIT | CONTROLLER, FOOT, HANDPIECE AND CORD | EBW | MAILLEFER INSTRUMENTS HOLDING SARL | 1046, 25572,00662 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |