FDA Adverse Event Malfunction Summary report: N

ALERE AFINION AS100 ANALYZER

MDR report key: 15499832 · Received September 28, 2022

Report

Report Number
3003045237-2022-00004
Event Type
Malfunction
Date Received
September 28, 2022
Date of Event
August 31, 2022
Report Date
May 25, 2023
Manufacturer
ABBOTT DIAGNOSTICS TECHNOLOGIES AS
Product Code
JQT
PMA / PMN Number
K180269
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION IS NOW CLOSED. CUSTOMER DID NOT RETURN INSTRUMENT / POWER PACK AS REQUESTED FOR INVESTIGATION. AS THE INSTRUMENT IS NOT AVAILABLE FOR, ASSIGNABLE CAUSE CANNOT BE DETERMINED. THE RISK OF HARM FROM FAILURES ASSOCIATED WITH THE REPORTED ISSUE IS ACCEPTABLE AND IS WITHIN THE PREDICTED OCCURRENCE RATE IN THE FMEA AND THE FAILURE RATE IS MONITORED BY TRACKING AND TRENDING. THE INVESTIGATION CONCLUDED THAT THE REPORTED EVENT IS NOT RELATED TO PRODUCT RESULTS, INADEQUATE INSTRUCTIONS FOR USE OR INACCURATE LABELS. THEREFORE, IT DOES NOT REPRESENT A POTENTIAL FOR PRODUCT DEFICIENCY PER INTERNAL PROCEDURES. THE CONCLUSION IS SUPPORTED BY THE MONTHLY COMPLAINT PRODUCT TREND REPORT .

Description of Event or Problem · 0

THE CUSTOMER EMAILED TECHNICAL SERVICES (TS) TO REPORT AN ISSUE WITH THEIR AFINION AS100 (1116054) , S/N (B)(4). DETAILS OF THE ISSUE: SPARKING AND MAKING GRINDING NOISE. THE FOLLOWING WAS OBSERVED WITH THE ANALYZER: VISIBLE SMOKE, BURNT SMELL, AND INSTRUMENT BEING TOO WARM/HOT TO TOUCH. NO INFORMATION REGARDING THE POWER CORD COULD BE PROVIDED. STATED THAT THE SPARKING OCCURRED IN THE HATCH AND NOT IN THE CABLE. NO DEATH OR SERIOUS INJURY REPORTED, AND NO IMPACT OR DELAY IN ANY TREATMENT. NOTED THAT NO ONE GOT HURT. CUSTOMER NO LONGER WANTED TO SWITCH ON THE ANALYZER TO THE REPORTED ALLEGATIONS PER THE CUSTOMERS PROCURED TENDER/DISTRIBUTOR AGREEMENT OF IMMEDIATE SWAPS FOR RENTAL UNITS WITHOUT TROUBLESHOOTING, A RETURN AND REPLACEMENT IS TO BE ORGANIZED.

Description of Event or Problem · 0

THE CUSTOMER EMAILED TECHNICAL SERVICES (TS) TO REPORT AN ISSUE WITH THEIR AFINION AS100 (1116054) , S/N (B)(6). DETAILS OF THE ISSUE: SPARKING AND MAKING GRINDING NOISE. THE FOLLOWING WAS OBSERVED WITH THE ANALYZER: - VISIBLE SMOKE - BURNT SMELL - INSTRUMENT BEING WARM/HOT TOO TOUCH NO INFORMATION REGARDING THE POWER CORD COULD BE PROVIDED. STATED THAT THE SPARKING OCCURRED IN THE HATCH AND NOT IN THE CABLE. NO DEATH OR SERIOUS INJURY REPORTED, AND NO IMPACT OR DELAY IN ANY TREATMENT. NOTED THAT NO ONE GOT HURT. CUSTOMER NO LONGER WANTED TO SWITCH ON THE ANALYZER TO THE REPORTED ALLEGATIONS PER THE CUSTOMERS PROCURED TENDER/DISTRIBUTOR AGREEMENT OF IMMEDIATE SWAPS FOR RENTAL UNITS WITHOUT TROUBLESHOOTING, A RETURN AND REPLACEMENT IS TO BE ORGANIZED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1633652 ALERE AFINION AS100 ANALYZER ANALYZER JQT ABBOTT DIAGNOSTICS TECHNOLOGIES AS 1116054 10200694

Patients

Seq Age Sex Outcome Treatment
1 Unknown