FDA Adverse Event Malfunction Summary report: N

VPV PROGRAMMER PLUS TRANSMIT JPN

MDR report key: 15495650 · Received September 27, 2022

Report

Report Number
3013886523-2022-00442
Event Type
Malfunction
Date Received
September 27, 2022
Report Date
November 8, 2022
Manufacturer
INTEGRA LIFESCIENCES MANSFIELD
Product Code
JXG
PMA / PMN Number
K061876
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THE VPV PROGRAMMER WAS RETURNED FOR EVALUATION: DHR - THE LOT MET SPECIFICATIONS WHEN RELEASED. FAILURE ANALYSIS - THE CUSTOMER¿S INDICATION OF ERROR IS CONFIRMED. THIS IS DUE TO A DEFECTIVE POWER MODULE AND PROGRAMMER BOX BOTTOM, WHICH MUST BE REPLACED. THE TRANSMITTER COIL FEET ARE BROKEN AND NEED REPLACEMENT. THE RESET, SAFETY AND FUNCTIONAL TEST MUST BE PERFORMED. REPAIR PERFORMED WITH RESET, SAFETY AND FUNCTIONAL TEST. ROOT CAUSE - THE ROOT CAUSES OF THE PROBLEMS WERE PROBABLY DUE TO NORMAL WEAR OUT (AS THE DEVICE IS OLDER THAN 15 YRS).

Description of Event or Problem · 0

A FACILITY REPORTED BURNING SMELL WHEN PLUGGING THE PROGRAMMER (ID 823192) IN AN ELECTRIC OUTLET AFTER CHANGING A SHUNT VALVE SETTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1951170 VPV PROGRAMMER PLUS TRANSMIT JPN VPV SYSTEM JXG INTEGRA LIFESCIENCES MANSFIELD

Patients

Seq Age Sex Outcome Treatment
1 Unknown