I-STAT TROPONIN (CTNI) CARTRIDGE
Report
- Report Number
- 2245578-2022-00140
- Event Type
- Malfunction
- Date Received
- September 27, 2022
- Date of Event
- September 20, 2022
- Report Date
- November 15, 2022
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- MMI
- UDI-DI
- 10054749000187
- PMA / PMN Number
- K031739
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 08-NOV-2022. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. RETAINED AND RETURNED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AK (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). NO DEFICIENCY HAS BEEN IDENTIFIED FOR CTNI CARTRIDGE LOT A22171B.
ON (B)(6) 2022, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED A DISCREPANT RESULT OF 6.37 NG/ML ON A 32 YEAR OLD MALE PATIENT WITH CHEST PAIN. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. I-STAT (B)(6) 2022 08:10 08:17 6.37 A, I-STAT (B)(6) 2022 08:10 08:50 0.00 A, I-STAT (B)(6) 2022 08:50 09:08 0.01 B. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THE INVESTIGATION IS UNDERWAY. PER I-STAT SYSTEM MANUAL: ART: 715595-00V, REPORTABLE RANGE: 0.00 - 50.00, REFERENCE RANGE: 0.00 - 0.03 (REPRESENTS THE 0 TO 97.5% RANGE OF RESULTS), REFERENCE RANGE: 0.00 - 0.08 (REPRESENTS THE 0 TO 99% RANGE OF RESULTS).