FDA Adverse Event Malfunction Summary report: N

ELECTROSURGICAL GENERATOR "ESG-400"

MDR report key: 15493769 · Received September 27, 2022

Report

Report Number
9610773-2022-00444
Event Type
Malfunction
Date Received
September 27, 2022
Date of Event
September 14, 2022
Report Date
September 27, 2022
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
GEI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE SUSPECT MEDICAL DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION/INVESTIGATION BUT TO OLYMPUS SHANGHAI (OSH), (RECEIVED AT OSH ON (B)(6) 2022). THE EVALUATION AT OSH CONFIRMED THE OCCURRENCE OF ERROR MESSAGES E433 AND E457 AND TRACED BOTH ERRORS BACK TO A DEFECT OF THE HVPS BOARD. THEREFORE, THIS EVENT/INCIDENT WAS ATTRIBUTED TO COMPONENT FAILURE. ANY ERROR MESSAGES THAT MAY APPEAR DURING OPERATION ARE TRIGGERED BY THE SAFETY SYSTEM OF THE ESG-400 AND COMMUNICATED VISUALLY AND ACOUSTICALLY TO THE USER. THEY ARE PART OF THE DEVICE'S OWN SECURITY CONCEPT. IN PARTICULARLY CRITICAL CASES, FURTHER USE OF THE DEVICE IS PREVENTED BY THE SECURITY SYSTEM UNTIL THE ERROR HAS BEEN CORRECTED. A MANUFACTURING AND QUALITY CONTROL REVIEW WAS PERFORMED FOR THE AFFECTED SERIAL NUMBER OF THE HF-GENERATOR WITHOUT SHOWING ANY ABNORMALITIES. THE CASE WILL BE CLOSED FROM OLYMPUS SIDE WITH NO FURTHER ACTIONS. THE REPORTED EVENT/INCIDENT WILL BE RECORDED FOR TRENDING AND SURVEILLANCE PURPOSES AND THE USER WILL BE INFORMED ABOUT THE INVESTIGATION RESULTS. PLEASE NOTE: THIS REPORT IS BEING SUBMITTED ALTHOUGH THE SUSPECT MEDICAL DEVICE IS NOT MARKETED IN THE USA. HOWEVER, A SIMILAR DEVICE IS MARKETED. MODEL # / CATALOG #: WB91051W; BRAND NAME: ELECTROSURGICAL GENERATOR "ESG-400"; COMMON DEVICE NAME: HF-GENERATORS; 510(K): K203682; PRODUCT CODE: GEI.

Description of Event or Problem · 0

THE LOANER ESG-400 ELECTROSURGICAL GENERATOR WAS REPORTED TO DISPLAY ERROR MESSAGE E433. NO PROCEDURE/PATIENT INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2258496 ELECTROSURGICAL GENERATOR "ESG-400" HF-GENERATORS GEI OLYMPUS WINTER & IBE GMBH WB91051C

Patients

Seq Age Sex Outcome Treatment
1 Unknown