BD SYRINGE LUER-LOK¿ TIP
Report
- Report Number
- 1213809-2022-00644
- Event Type
- Malfunction
- Date Received
- September 27, 2022
- Date of Event
- August 11, 2022
- Report Date
- October 19, 2022
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- FMF
- UDI-DI
- 00382903096053
- PMA / PMN Number
- K110771
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVALUTION?: YES. D10: RETURNED TO MANUFACTURER ON: 10-OCT-2022 H6: INVESTIGATION SUMMARY ONE SAMPLE WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, A SEVERE INDENTATION OBSERVED ON THE TOP OF THE BARREL COLLAR. POTENTIAL ROOT CAUSE FOR THE DAMAGED COLLAR DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS. THIS CONDITION IS OCCURRING BELOW THEIR EXPECTED FREQUENCY SO NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 1200434. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.
IT WAS REPORTED THAT THE BD SYRINGE LUER-LOK¿ TIP ARE DENTED/CHIPPED OFTEN CAUSING THE SYRINGE TO LEAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE TIPS OF THE SYRINGES ARE DENTED/CHIPPED AND OFTEN CAUSING THE SYRINGES TO LEAK. WE HAVE SEEN 10-15 SYRINGES WITH THIS DEFECT SO FAR.
IT WAS REPORTED THAT THE BD SYRINGE LUER-LOK¿ TIP ARE DENTED/CHIPPED OFTEN CAUSING THE SYRINGE TO LEAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE TIPS OF THE SYRINGES ARE DENTED/CHIPPED AND OFTEN CAUSING THE SYRINGES TO LEAK. WE HAVE SEEN 10-15 SYRINGES WITH THIS DEFECT SO FAR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2078417 | BD SYRINGE LUER-LOK¿ TIP | PISTON SYRINGE | FMF | BECTON DICKINSON MEDICAL SYSTEMS | 1200434 | 00382903096053 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |