BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM
Report
- Report Number
- 3014704491-2022-00445
- Event Type
- Malfunction
- Date Received
- September 27, 2022
- Date of Event
- September 5, 2022
- Report Date
- October 2, 2022
- Manufacturer
- BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
- Product Code
- FOZ
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H6: INVESTIGATION SUMMARY: IN RESPONSE TO THE EVENT REPORTED A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 1111432. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. ALTHOUGH PHOTOS WERE SUBMITTED FOR EVALUATION, THE AFFECTED SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM DISCONNECTED AND NOT BROKEN IN THE BODY. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: DURING THE PUNCTURE, IT WAS FOUND THAT THE FRONT END OF THE CATHETER TUBE WAS COMPLETELY DISCONNECTED AND NOT BROKEN IN THE BODY.
IT WAS REPORTED THAT THE BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM DISCONNECTED AND NOT BROKEN IN THE BODY. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: DURING THE PUNCTURE, IT WAS FOUND THAT THE FRONT END OF THE CATHETER TUBE WAS COMPLETELY DISCONNECTED AND NOT BROKEN IN THE BODY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2304091 | BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM | INTRAVASCULAR CATHETER | FOZ | BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU | 1111432 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |