IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM
Report
- Report Number
- 0002023141-2022-02450
- Event Type
- Injury
- Date Received
- September 27, 2022
- Date of Event
- June 26, 2022
- Report Date
- February 3, 2023
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024020047
- PMA / PMN Number
- K061410
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- 003
Narratives
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). 220825-11-L.Y. ADDITIONAL PMA/510(K) NUMBER ¿ K011028/K013227. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR INFECTION EVENTS IDENTIFYING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO THE WIDE RANGE OF EXTERNAL (NON-DESIGN/ NON-MANUFACTURING RELATED) FACTORS POTENTIALLY IMPACTING THE STERILITY OF THE IMPLANT AND ITS ENVIRONMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE INFECTION EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.
THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: H6: TYPE OF INVESTIGATION CODE WAS ADDED: 3331. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1240553). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. STERILIZATION RECORD (OP160) WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1240553) FOR SIMILAR EVENTS AND ONE OTHER COMPLAINT WAS IDENTIFIED. INVESTIGATION HAS IDENTIFIED THAT THE MOST LIKELY PROBABLE CAUSES ARE RELATED TO MEDICAL CONDITIONS/PATIENT HABITS AND SURGICAL TECHNIQUE. AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMMER BIOMET CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATION, NO MALFUNCTION OCCURRED UPON INVESTIGATION. THE REPORTED EVENT REMAINS NON-VERIFIABLE AS IT IS A MEDICAL CONDITION.
IT WAS REPORTED THAT THE PATIENT FELT PAIN AT THE IMPLANT SITE AT TOOTH LOCATION #34. THE PATIENT CAME TO THE CLINIC FOR EXAMINATION AND IMPLANT INFECTION WAS FOUND. THE IMPLANT WAS THEN REMOVED FOR ANTI-INFLAMMATORY TREATMENT. SYMPTOMS AS A RESULT OF THE EVENT: PAIN AND INFLAMMATION.
NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2045361 | IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | TSVWB10 | 1240553 | 00889024020047 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR | Male | Required Intervention |