FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK¿ SYRINGE

MDR report key: 15486856 · Received September 26, 2022

Report

Report Number
9614033-2022-00076
Event Type
Malfunction
Date Received
September 26, 2022
Date of Event
August 10, 2022
Report Date
October 3, 2022
Manufacturer
BECTON DICKINSON DE MEXICO
Product Code
FMF
UDI-DI
07501073025530
PMA / PMN Number
K151766
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: B5. IT WAS REPORTED THAT 2 BD PLASTIPAK¿ SYRINGE HAD DIFFICULT PLUNGER MOVEMENT. THIS COMPLAINT WAS CREATED TO CAPTURE THE 2ND OF 2 RELATED INCIDENTS.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: TWO PHOTOS RECEIVED BY OUR QUALITY TEAM FOR INVESTIGATION. UPON VISUAL INSPECTION, PRODUCT LABEL INFORMATION CATALOG 302553, BATCH NUMBER 7230949 IS DISPLAYED, THE PRODUCT IS EXPIRED. NO OTHER DEFECTS OR ISSUES CAN BE OBSERVED, THEREFORE THIS INCIDENT IS NOT CONFIRMED. THE SAMPLES HAVE ALREADY PASSED THEIR EXPIRATION DATE, THEREFORE BD DOES NOT GUARANTEE THAT THE PRODUCT PRESENTS ANY ALTERATION OR MEETING SPECIFICATION. BASED ON THE QUALITY TEAM'S INVESTIGATION, WE ARE NOT ABLE TO IDENTIFY A ROOT CAUSE RELATED TO OUR MANUFACTURING PROCESS AT THIS TIME.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT 2 BD PLASTIPAK¿ SYRINGE HAD DIFFICULT PLUNGER MOVEMENT. THIS COMPLAINT WAS CREATED TO CAPTURE THE 2ND OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM SPANISH: "CYLINDER MADE OF MEDICAL GRADE MATERIAL, WHICH FITS PERFECTLY INSIDE THE BARREL AND IS EASILY MOVED MANUALLY BOTH INSIDE IT. ALTHOUGH THE PLUNGER MOVES, IT DOES NOT MOVE EASILY. IT MUST BE TAKEN INTO ACCOUNT THAT, FOR A REPETITIVE ACTION, THE USER COULD HAVE PAIN AND/OR INJURY, IN ADDITION TO INCURRING AN ADVERSE EFFECT ON THE USE WITH THE PATIENT."

Description of Event or Problem · 0

IT WAS REPORTED THAT 2 BD PLASTIPAK¿ SYRINGE HAD DIFFICULT PLUNGER MOVEMENT. THIS COMPLAINT WAS CREATED TO CAPTURE THE 2ND OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM SPANISH: "CYLINDER MADE OF MEDICAL GRADE MATERIAL, WHICH FITS PERFECTLY INSIDE THE BARREL AND IS EASILY MOVED MANUALLY BOTH INSIDE IT. ALTHOUGH THE PLUNGER MOVES, IT DOES NOT MOVE EASILY. IT MUST BE TAKEN INTO ACCOUNT THAT, FOR A REPETITIVE ACTION, THE USER COULD HAVE PAIN AND/OR INJURY, IN ADDITION TO INCURRING AN ADVERSE EFFECT ON THE USE WITH THE PATIENT."

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD PLASTIPAK¿ SYRINGE HAD DIFFICULT PLUNGER MOVEMENT. THIS COMPLAINT WAS CREATED TO CAPTURE THE 2ND OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM SPANISH: "CYLINDER MADE OF MEDICAL GRADE MATERIAL, WHICH FITS PERFECTLY INSIDE THE BARREL AND IS EASILY MOVED MANUALLY BOTH INSIDE IT. ALTHOUGH THE PLUNGER MOVES, IT DOES NOT MOVE EASILY. IT MUST BE TAKEN INTO ACCOUNT THAT, FOR A REPETITIVE ACTION, THE USER COULD HAVE PAIN AND/OR INJURY, IN ADDITION TO INCURRING AN ADVERSE EFFECT ON THE USE WITH THE PATIENT."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1768290 BD PLASTIPAK¿ SYRINGE PISTON SYRINGE FMF BECTON DICKINSON DE MEXICO 7230949 07501073025530

Patients

Seq Age Sex Outcome Treatment
1 Unknown