L-CATH PICC S/L 26GA (1.9F) 0.60MM X 30CM
Report
- Report Number
- 0001625425-2022-01100
- Event Type
- Malfunction
- Date Received
- September 26, 2022
- Date of Event
- July 19, 2021
- Report Date
- October 22, 2022
- Manufacturer
- ARGON MEDICAL DEVICES
- Product Code
- LJS
- UDI-DI
- 00886333209934
- PMA / PMN Number
- K091670
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
A REVIEW OF THE DHR AND INSPECTION RECORDS WAS CONDUCTED, AND NO SIMILAR CONCERNS WERE FOUND. ONE OPENED CATHETER WAS RETURNED FOR REVIEW. VISUAL INSPECTION CONFIRMED A CRACK IN THE LUER THAT WOULD RESULT IN LEAKAGE. SEVERAL COMPLAINTS FOR THIS PART NUMBER HAVE PREVIOUSLY BEEN RECEIVED REGARDING A CRACKED HUB RESULTING IN LEAKAGE, AND CAPA 2021-039 WAS INITIATED TO ADDRESS THIS ISSUE. THE CAPA IS CURRENTLY IN THE IMPLEMENTATION PHASE AND WILL EVALUATE THE CORRECTIVE ACTION IMPLEMENTATION FOR EFFECTIVENESS TO PREVENT A RECURRENCE OF THIS ISSUE.
THE SAMPLE IS INDICATED AS AVAILABLE FOR RETURN. AS OF THE DATE OF THIS REPORT, THE SAMPLE HAS NOT BEEN RETURNED. A FOLLOW-UP REPORT WILL BE PROVIDED ONCE THE DEVICE HAS BEEN RECEIVED AND REVIEWED.
DATE PICC PLACED. (B)(6) 19 @1415. DATE OF PRODUCT EXPERIENCE. (B)(6) 22 @ 0100. SKU# 384539. LOT# 11410069. DESCRIBE EVENT. PLACED MIDLINE WITHOUT DIFFICULTY. THE NIGHT SHIFT NURSE NOTICED THE DRESSING WAS DAMP. UPON INVESTIGATION SHE FOUND THAT THERE WAS A CRACK IN THE HUB ANY HARM TO PATIENT? NOT DOCUMENTED IN INCIDENT REPORT FILED. IF SO, PLEASE DESCRIBE. WAS A NEW PIV PLACED FOR COMPLETION OF PRESCRIBED THERAPY? WAS A NEW PICC PLACED FOR COMPLETION OF PRESCRIBED THERAPY? YES . IS PRODUCT AVAILABLE FOR RETURN? YES.
DATE PICC PLACED. 7/19 @1415 DATE OF PRODUCT EXPERIENCE. (B)(6) 2022 @ 0100. LOT# 11410069. DESCRIBE EVENT. PLACED MIDLINE WITHOUT DIFFICULTY. THE NIGHT SHIFT NURSE NOTICED THE DRESSING WAS DAMP. UPON INVESTIGATION SHE FOUND THAT THERE WAS A CRACK IN THE HUB ANY HARM TO PATIENT? NOT DOCUMENTED IN INCIDENT REPORT FILED. IF SO, PLEASE DESCRIBE. WAS A NEW PIV PLACED FOR COMPLETION OF PRESCRIBED THERAPY? WAS A NEW PICC PLACED FOR COMPLETION OF PRESCRIBED THERAPY? YES. IS PRODUCT AVAILABLE FOR RETURN? YES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2930463 | L-CATH PICC S/L 26GA (1.9F) 0.60MM X 30CM | L-CATH PICC | LJS | ARGON MEDICAL DEVICES | 384539 | 11410069 | 00886333209934 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |