PENTAX

MDR report key: 15482135 · Received September 25, 2022

Report

Report Number
9610877-2022-58162
Event Type
Malfunction
Date Received
September 25, 2022
Date of Event
September 13, 2022
Report Date
September 26, 2022
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL VNL-1570STK IS AVAILABLE IN THE USA WITH A 510K NUMBER K162151. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD MODULE BLACKOUT. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD MODULE. IN ADDITION, WE CONFIRMED THAT THE REMOTE CONTROL BUTTONS PERFORATED, THE INSERTION FLEXIBLE TUBE (IFT) BUCKLED, THE LIGHT GUIDE CABLE FOR CONTROL BODY BUCKLED, THE LIGHT GUIDE CABLE FOR PRONG BUCKLED, THE LIGHT GUIDE CABLE CUT, THE ROOT BRACE RUBBER CUT, AND THE U/D KNOB LOOSE; HOWEVER, THEY ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT ).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2076823 PENTAX VIDEO NASO PHARYNGO LARYNGOSTROBOSCOPE EOB HOYA CORPORATION PENTAX TOKYO OFFICE VNL-1070STK

Patients

Seq Age Sex Outcome Treatment
1 Unknown