FDA Adverse Event Malfunction Summary report: N

ALM PRISMATIC

MDR report key: 154797 · Received March 5, 1998

Report

Report Number
2027060-1998-00001
Event Type
Malfunction
Date Received
March 5, 1998
Date of Event
February 5, 1998
Report Date
March 5, 1998
Manufacturer
ALM SA PARC DE LIMERE
Product Code
FQP
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

IT WAS REPORTED THAT ON 2/5/98, A SMALL AREA OF PAINT FROM THE ALM PRC LIGHT CHIPPED AND FELL INTO THE SURGICAL FIELD. THE HOSP REQUESTED THAT CO'S PROJECT AND TECHNICAL SPECIALISTS MEET WITH THE HOSP TO INSPECT THE LIGHT. THIS TOOK PLACE ON 2/7/98. UPON INSPECTION, IT WAS NOTED THAT THE PAINT HAD CHIPPED FROM AN ISOLATED AREA OF THE LIGHTHEAD SUPPORT ARM SYSTEM WHICH SEEMED TO HAVE BEEN SUBJECTED TO CONTINUOUS IMPACT OVER TIME. THIS SURGICAL LIGHT IS APPROX 4 YRS, 10 MONTHS OLD. A REPLACEMENT SUPPORT ARM WAS SENT FROM CO'S FACILITY ON 2/13/98 TO ELIMINATE FURTHER PROBLEMS. CO SUGGESTED THAT THE HOSP REFURBISH THEIR OLDER LIGHTING SYSTEMS AND UNDERGO FURTHER IN SERVICING ON PROPER POSITIONING OF SURGICAL LIGHTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ALM PRISMATIC SURGICAL LIGHTING FQP ALM SA PARC DE LIMERE 555012 NA

Patients

Seq Age Sex Outcome Treatment