FDA Adverse Event Malfunction Summary report: N

ETHMOID FORCEPS STRAIGHT

MDR report key: 1547910 · Received November 20, 2009

Report

Report Number
1418479-2009-00017
Event Type
Malfunction
Date Received
November 20, 2009
Date of Event
October 7, 2009
Report Date
November 20, 2009
Manufacturer
RICHARD WOLF GMBH
Product Code
JXX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: 8211.09 - ETHMOID FORCEPS STRAIGHT - LOT #: UNK. THE ETHMOID FORCEPS WERE INSPECTED AND BROKEN JAW WAS CONFIRMED. THE BROKEN JAW WAS NOT INCLUDED IN THE RETURNED PACKAGE. WE DO NOT KNOW IF IT WAS DISCARDED WITHIN OUR FACILITY, OR AT THE USER FACILITY. THE ETHMOID FORCEPS IS, FROM APPEARANCE, 10 PLUS YEARS OLD. THE ETHMOID FORCEPS ARE DESIGNED TO GRASP, NOT CUT TISSUE. THE TISSUE IS REMOVED BY GRASPING AND PULLING IT GENTLY AWAY FROM THE MUSCOSA OR BONE. CAUSE: USE AND HANDLING WITH EXCESS FORCE APPLIED TO THE JAW DURING USE.

Description of Event or Problem · 1

DURING A SINUS PROCEDURE (LANDMARX IMAGE GUIDED FESS), THE JAW BROKE. THE JAW PIECE AND FORCEPS WERE COLLECTED FOR RETURN TO MANUFACTURER. THERE WAS NO PT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ETHMOID FORCEPS STRAIGHT ETHMOID FORCEPS STRAIGHT JXX RICHARD WOLF GMBH 8211.09 UNK

Patients

Seq Age Sex Outcome Treatment
1 28 YR