FDA Adverse Event
Malfunction
Summary report: N
ETHMOID FORCEPS STRAIGHT
MDR report key: 1547910
·
Received November 20, 2009
Report
- Report Number
- 1418479-2009-00017
- Event Type
- Malfunction
- Date Received
- November 20, 2009
- Date of Event
- October 7, 2009
- Report Date
- November 20, 2009
- Manufacturer
- RICHARD WOLF GMBH
- Product Code
- JXX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: 8211.09 - ETHMOID FORCEPS STRAIGHT - LOT #: UNK. THE ETHMOID FORCEPS WERE INSPECTED AND BROKEN JAW WAS CONFIRMED. THE BROKEN JAW WAS NOT INCLUDED IN THE RETURNED PACKAGE. WE DO NOT KNOW IF IT WAS DISCARDED WITHIN OUR FACILITY, OR AT THE USER FACILITY. THE ETHMOID FORCEPS IS, FROM APPEARANCE, 10 PLUS YEARS OLD. THE ETHMOID FORCEPS ARE DESIGNED TO GRASP, NOT CUT TISSUE. THE TISSUE IS REMOVED BY GRASPING AND PULLING IT GENTLY AWAY FROM THE MUSCOSA OR BONE. CAUSE: USE AND HANDLING WITH EXCESS FORCE APPLIED TO THE JAW DURING USE.
Description of Event or Problem · 1
DURING A SINUS PROCEDURE (LANDMARX IMAGE GUIDED FESS), THE JAW BROKE. THE JAW PIECE AND FORCEPS WERE COLLECTED FOR RETURN TO MANUFACTURER. THERE WAS NO PT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ETHMOID FORCEPS STRAIGHT | ETHMOID FORCEPS STRAIGHT | JXX | RICHARD WOLF GMBH | 8211.09 | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR |