FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 ANTIGEN SELF TEST

MDR report key: 15478379 · Received September 24, 2022

Report

Report Number
1221359-2022-04499
Event Type
Malfunction
Date Received
September 24, 2022
Date of Event
September 13, 2022
Report Date
September 23, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO PROVIDE THE INITIAL REPORT AND INVESTIGATION CONCLUSION. TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4), INC. ON RETAINED KIT LOT 173451 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160 / LOT 173451 AND DEVICE PART NUMBER 195-430H / LOT 170827. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED ASFALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 173451 SHOWED THAT THE COMPLAINT RATE IS (B)(4). IN CONCLUSION, ABBOTT DIAGNOSTICS (B)(4) WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SPECIFIC PATIENT SAMPLE.

Description of Event or Problem · 0

THE CONSUMER REPORTED AN UNCONFIRMED FALSE NEGATIVE RESULTS WITH THE BINAX NOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON (B)(6) 2022. THE CONSUMER PERFORMED AN ALTERNATE TESTING ON (B)(6) 2022 USING AN IHEALTH TEST THAT GENERATED A POSITIVE RESULT. CONFIRMATION TESTING WAS NOT PERFORMED. THE PATIENT WAS SYMPTOMATIC (SLIGHT HEADACHE, FEELING "FIZZY", SORE THROAT).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
127097 BINAXNOW COVID-19 ANTIGEN SELF TEST CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 173451 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 81 YR Female