FDA Adverse Event Malfunction Summary report: N

X SERIES

MDR report key: 15478298 · Received September 24, 2022

Report

Report Number
1220908-2022-03573
Event Type
Malfunction
Date Received
September 24, 2022
Report Date
September 5, 2022
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
P160022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ZOLL MEDICAL CORPORATION HAS NOT RECEIVED THE DEVICE FOR EVALUATION AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO A ZOLL APPROVED SERVICE PROVIDER FOR EVALUATION. THE CUSTOMER'S REPORT WAS OBSERVED DURING REVIEW OF THE DEVICE DATA LOGS. HOWEVER, THE DEVICE WAS PUT THROUGH EXTENSIVE TESTING INCLUDING BENCH HANDLING AND FULL FUNCTIONAL STRESS TESTING WITHOUT DUPLICATING THE REPORT. THE MILTIFUNCTION CABLE (MFC) AND MFC RECEPTACLE WERE REPLACED AS A PRECAUTION. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. THE MFC AND MFC RECEPTACLE WERE RETURNED TO ZOLL MEDICAL CORPORATION AND SCRAPPED. ANALYSIS OF REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 0

COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED A "ECG MONITORING FAILURE" MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1604168 X SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION X SERIES NA

Patients

Seq Age Sex Outcome Treatment
1 Unknown