FDA Adverse Event Malfunction Summary report: N

BD PRESET¿ ECLIPSE¿

MDR report key: 15474404 · Received September 23, 2022

Report

Report Number
9617032-2022-00945
Event Type
Malfunction
Date Received
September 23, 2022
Date of Event
September 16, 2022
Report Date
October 6, 2022
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
PMA / PMN Number
K022426
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL LOT NUMBER IS AS FOLLOWS: D.4. MEDICAL DEVICE LOT #: 2182401. D.4. MEDICAL DEVICE EXPIRATION DATE: 2024-06-30. H.4. DEVICE MANUFACTURE DATE: 2022-07-01. D9: DEVICE AVAILABLE FOR EVAL: YES. D9: RETURNED TO MANUFACTURER ON: 2022-10-05. H.6. INVESTIGATION SUMMARY: "MATERIAL #: 364390. LOT/BATCH #: 2182401 & 2172749. BD RECEIVED 6 SAMPLES AND 1 PHOTO FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND THE INDICATED FAILURE MODE FOR LOOSE CAP CAUSING LEAKAGE WAS OBSERVED. ADDITIONALLY, THE CUSTOMER SAMPLES ALONG WITH 10 RETENTION SAMPLES FROM BD INVENTORY, WERE EVALUATED BY FUNCTIONAL TESTING, EACH DRAWN WITH WATER AND ALL HEMOGARD CAPS WERE TIGHTENED TO THE SYRINGE. THE ISSUE OF LOOSE CAP WAS NOT OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE LOOSE CAP. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD PRESET¿ ECLIPSE¿, THE DEVICE EXPERIENCED FAILURE OF THE PRODUCT TO CONTAIN BLOOD. THE AMOUNT OF OCCURRENCE IS UNCLEAR BUT AN ESTIMATED AMOUNT OF 500 PRODUCTS WERE NOTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: SYRINGES WITH IMPROPERLY CAPPED SAMPLES AND SPILLED SAMPLES ARE RECEIVED IN THE LABORATORY. THEY HAVE RECEIVED IN THE LABORATORY SAMPLES FROM ALL OVER THE HOSPITAL THAT HAD SPILLED IN TRANSPORT BECAUSE THE CAPS DID NOT CLOSE PROPERLY. THEY HAVE SEEN THAT IT SEEMS THAT THE CAPS THEMSELVES DO NOT HAVE THE THREAD PROPERLY MADE TO CONNECT PROPERLY WITH THE LUER LOK CONE OF THE SYRINGES. THEY ARE NOT CLEAR ABOUT THE AFFECTED LOT BECAUSE FROM THE WAREHOUSE THEY ARE DISTRIBUTED THROUGHOUT THE HOSPITAL AND FROM THE SERVICES WHERE THEY TAKE THE SAMPLES THEY HAVE NOT NOTICED THAT THE STOPPER DID NOT CONNECT CORRECTLY, IT WAS IN THE LABORATORY WHEN THEY RECEIVED THEM WHEN THEY DETECTED THAT THE SAMPLE HAD SPILLED DURING TRANSPORT AND THEY HAVE NO WAY OF KNOWING WHICH WAS THE LOT OF THE PRODUCT.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD PRESET¿ ECLIPSE¿, THE DEVICE EXPERIENCED FAILURE OF THE PRODUCT TO CONTAIN BLOOD. THE AMOUNT OF OCCURRENCE IS UNCLEAR BUT AN ESTIMATED AMOUNT OF 500 PRODUCTS WERE NOTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: SYRINGES WITH IMPROPERLY CAPPED SAMPLES AND SPILLED SAMPLES ARE RECEIVED IN THE LABORATORY. THEY HAVE RECEIVED IN THE LABORATORY SAMPLES FROM ALL OVER THE HOSPITAL THAT HAD SPILLED IN TRANSPORT BECAUSE THE CAPS DID NOT CLOSE PROPERLY. THEY HAVE SEEN THAT IT SEEMS THAT THE CAPS THEMSELVES DO NOT HAVE THE THREAD PROPERLY MADE TO CONNECT PROPERLY WITH THE LUER LOK CONE OF THE SYRINGES. THEY ARE NOT CLEAR ABOUT THE AFFECTED LOT BECAUSE FROM THE WAREHOUSE THEY ARE DISTRIBUTED THROUGHOUT THE HOSPITAL AND FROM THE SERVICES WHERE THEY TAKE THE SAMPLES THEY HAVE NOT NOTICED THAT THE STOPPER DID NOT CONNECT CORRECTLY, IT WAS IN THE LABORATORY WHEN THEY RECEIVED THEM WHEN THEY DETECTED THAT THE SAMPLE HAD SPILLED DURING TRANSPORT AND THEY HAVE NO WAY OF KNOWING WHICH WAS THE LOT OF THE PRODUCT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2828414 BD PRESET¿ ECLIPSE¿ BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 2172749

Patients

Seq Age Sex Outcome Treatment
1 Unknown