FDA Adverse Event Malfunction Summary report: N

BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC)

MDR report key: 15471697 · Received September 23, 2022

Report

Report Number
3008352382-2022-00006
Event Type
Malfunction
Date Received
September 23, 2022
Date of Event
September 14, 2022
Report Date
February 15, 2023
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
MDB
UDI-DI
00382904420215
PMA / PMN Number
K123903
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6 INVESTIGATION SUMMARY: CATALOG 442021. BATCH NO. 2070284. CUSTOMER REPORTED A PAENIBACILLUS CONTAMINATION ISSUE WHILE USING BACTEC PRODUCT. ONE LIST OF BARCODE SEQUENCES WAS RECEIVED. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH BACTEC PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED, AND BATCH HAS BEEN PREVIOUSLY INVESTIGATED FOR THE REPORTED DEFECT. SEVERAL EVALUATIONS WERE PERFORMED TO CONCLUDE THAT THE BIO-CONTAMINANT PAENIBACILLUS DID NOT ORIGINATED FROM BD MANUFACTURING SITE. THE AUTOCLAVE PROCESS WAS EVALUATED. THE MOIST HEAT AUTOCLAVE IS A PROCESS THAT USES MOIST HEAT AS THE LETHAL AGENT TO RENDER LIQUID AND POROUS/HARD GOODS ITEMS FREE FROM VIABLE MICROORGANISM AND RELIES ON DIRECT CONTACT WITH SATURATED STEAM, WHICH IS WATER VAPOR IN EQUILIBRIUM WITH LIQUID WATER. THE AUTOCLAVE PROCESS IS ACCOMPLISHED BY THE HEAT TRANSFER FROM THE STEAM TO THE LOAD AND BY HYDRATING EFFECT OF THE RESULTANT CONDENSATE THAT FORMS DUE TO THE CHANGE IN STATE FROM VAPOR TO THE LOWER ENERGY LIQUID STATE. THIS PHASE CHANGE REQUIRES THE TRANSFER OF THE LATENT HEAT OF THE STEAM TO THE SURROUNDINGS, THUS HEATING THE AUTOCLAVE AND ITS LOAD. BD CAYEY AUTOCLAVE THE PRODUCTS VIA MOIST HEAT. INITIAL VALIDATION HAD BEEN SUCCESSFULLY COMPLETED AND THE REQUIRED REQUALIFICATION ASSESSMENT HAS BEEN PERFORMED ACCORDINGLY MAINTAINING THE ACTIVE AND VALID AUTOCLAVING CYCLE. ENVIRONMENTAL SAMPLES WERE EVALUATED FROM BD MANUFACTURING SITE FOR IDENTIFICATION OF BACTERIAL SPECIES PRESENT AT DIFFERENT LOCATIONS WITHIN THE MANUFACTURING PLANT. THE PURPOSE OF IDENTIFYING THESE ISOLATES WAS TO DETERMINE IF PAENIBACILLUS CONTAMINATION ORIGINATED FROM THIS LOCATION. AFTER PERFORMING MALDI TESTING NONE OF THE ENVIRONMENTAL SAMPLES COLLECTED FROM DIFFERENT LOCATIONS WITHIN THE MANUFACTURING PLANT IN BD CAYEY WERE IDENTIFIED AS PAENIBACILLUS. THEREFORE, THE BIO-CONTAMINANT PAENIBACILLUS LIKELY DID NOT ORIGINATE FROM THE BD CAYEY, AND THERE IS NO EVIDENCE TO INDICATE THAT IT WOULD IT HAVE SURVIVED THE AUTOCLAVING PROCESS OF BACTEC BOTTLES. PRODUCT INSERT WARNINGS AND PRECAUTIONS SECTIONS STATES THAT PRIOR TO USE, EACH VIAL SHOULD BE EXAMINED FOR EVIDENCE OF CONTAMINATION SUCH AS CLOUDINESS, BULGING OR DEPRESSES SEPTUM, OR LEAKAGE. VIALS SHOWING EVIDENCE OF CONTAMINATION SHOULD NOT BE USED. ALSO PRIOR TO USE, THE USER SHOULD EXAMINE THE VIAL FOR EVIDENCE OF DAMAGE OR DETERIORATION. VIALS DISPLAYING TURBIDITY, CONTAMINATION, OR DISCOLORATION (DARKENING) SHOULD NOT BE USED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS.

Additional Manufacturer Narrative · 0

LOT#: UNKNOWN. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. INITIAL REPORTER EMAIL: (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN.

Description of Event or Problem · 0

IT WAS REPORTED BY THE CUSTOMER THAT BOTTLES HAD CONTAMINATION - BIOLOGICAL CONTAMINATION - BACTERIAL, FUNGAL, NON-VIABLE, DUE TO ISOLATION OF PAENIBACILLUS URINALIS GROUP FROM BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC). THERE WAS ONE OCCURRENCE IN THIS EVENT WITH NO PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE CUSTOMER: "CUSTOMER ALLEGES CONTAMINATED BOTTLES DUE TO ISOLATION OF PAENIBACILLUS URINALIS GROUP FROM SEVERAL BOTTLES.".

Description of Event or Problem · 0

IT WAS REPORTED BY THE CUSTOMER THAT BOTTLES HAD CONTAMINATION - BIOLOGICAL CONTAMINATION - BACTERIAL, FUNGAL, NON-VIABLE, DUE TO ISOLATION OF PAENIBACILLUS URINALIS GROUP FROM BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC). THERE WAS ONE OCCURRENCE IN THIS EVENT WITH NO PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE CUSTOMER: "CUSTOMER ALLEGES CONTAMINATED BOTTLES DUE TO ISOLATION OF PAENIBACILLUS URINALIS GROUP FROM SEVERAL BOTTLES."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2828745 BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) SYSTEM, BLOOD CULTURING MDB BECTON DICKINSON CARIBE LTD. 442021 UNKNOWN 00382904420215

Patients

Seq Age Sex Outcome Treatment
1 Unknown