SIX WIRE LITHOTRIPTOR BASKET, 30 MM LENGTH, 2.6 MM DIAMETER

MDR report key: 154703 · Received February 10, 1998

Report

Report Number
9680809-1998-90001
Date Received
February 10, 1998
Date of Event
February 2, 1998
Manufacturer
GIP MEDIZINTECHNIK GMBH
Product Code
LQC
Report Source
Manufacturer report
Reporter Occupation
INVALID DATA

Narratives

Description of Event or Problem · 1

"DURING AN ERCP PROCEDURE A STONE WAS FOUND IN A PTS COMMON BILE DUCT. AFTER THE STONE WAS CAPTURED WITHIN THE BASKET AND THE LITHOTRIPTOR HANDLE WAS CRANKED IN AN ATTEMPT TO CURSH THE STONE, THE LITHOTRIPTOR WIRE SNAPPED. WHEN THE METAL SPIRAL WAS WITHDRAWN FROM THE DUODENOSCOPE, A TWELVE TO EIGHTEEN INCH LENGTH OF FRAYED WIRE WAS LEFT PROTRUDING FROM THE ENDOSCOPE'S CHANNEL INLET. THE WIRE HAD SNAPPED AND WAS COMPLETELY SEVERED AT APPROXIMATELY 73 INCHES FROM DISTAL END. SUBSEQUENTLY, AN ATTEMPT WAS MADE TO FREE THE STONE FROM THE BASKET. HOWEVER, WHEN THIS WAS UNSUCCESSFUL THE PT HAD TO HAVE OPEN SURGERY TO REMOVE THE WIREBASKET AS WELL AS SEVERAL STONES RETAINED IN THE GALL BLADDER."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIX WIRE LITHOTRIPTOR BASKET, 30 MM LENGTH, 2.6 MM DIAMETER SIX WIRE LITHOTRIPTOR BASKET, 30 MM LENGTH, 2.6 MM DIAMETER LQC GIP MEDIZINTECHNIK GMBH * 96111871