FDA Adverse Event Malfunction Summary report: N

CADD LEGACY PLUS PUMP

MDR report key: 15470139 · Received September 23, 2022

Report

Report Number
3012307300-2022-20190
Event Type
Malfunction
Date Received
September 23, 2022
Report Date
October 19, 2022
Manufacturer
ST PAUL
Product Code
FRN
PMA / PMN Number
K982838
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

UNIQUE IDENTIFIER (UDI) AND PROTOCOL # ARE UNKNOWN. NO INFORMATION HAS BEEN PROVIDED TO DATE. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

DEVICE AVAILABLE FOR EVALUATION: UPDATED. H3: DEVICE EVALUATED BY MANUFACTURER: UPDATED. H6: EVENT PROBLEM AND EVALUATION CODES: UPDATED. H10: DEVICE EVALUATION: THE DEVICE WAS RETURNED TO MANUFACTURING FOR INVESTIGATION. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE PERFORMED. VISUAL INSPECTION OF THE RETURNED DEVICE CONFIRMED THAT THE TAMPER SEALS AND PHYSICAL CONDITION WERE ALL INTACT UPON RECEIPT. ERROR CODE WAS FOUND IN THE EHL (EVENT HISTORY LOG). THE CUSTOMER REPORTED PROBLEM WAS NOT DUPLICATED DURING THE INVESTIGATION. THE ROOT CAUSE OF THE REPORTED EVENT IS UNKNOWN AS NO FAULT WAS FOUND. PREVENTIVELY, REPLACE THE DOWNSTREAM SENSOR. A DEVICE HISTORY RECORD DHR REVIEW WAS NOT CONDUCTED, BASED UPON REVIEW OF THE INFORMATION PROVIDED BY THE CUSTOMER, AS IT DOES NOT INDICATE A PROBLEM WITH THE INITIAL MANUFACTURE OR PRIOR REPAIR OF THE DEVICE., CORRECTED DATA: G5: CORRECTION: PREMARKET 510K NUMBER: K982838.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE DEVICE HAD A NO DISPOSABLE ALARM. THERE HAS BEEN NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
553499 CADD LEGACY PLUS PUMP PUMP, INFUSION FRN ST PAUL 6500

Patients

Seq Age Sex Outcome Treatment
1 Unknown