FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 15468277 · Received September 22, 2022

Report

Report Number
9610877-2022-58147
Event Type
Malfunction
Date Received
September 22, 2022
Date of Event
September 7, 2022
Report Date
September 22, 2022
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EG-2990K IS AVAILABLE IN THE USA WITH A 510K NUMBER K131902. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE OPERATION CHANNEL BLOCKED. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE INSUFFICIENT REPROCESSING AT THE FACILITY. BASED ON THE TECHNICAL REPORT ""HR-RPT-0588(CHANNEL)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. OPERATION/SUCTION CHANNEL CLOGGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
565342 PENTAX VIDEO GASTROSCOPE FDS HOYA CORPORATION PENTAX TOKYO OFFICE EG-290KP

Patients

Seq Age Sex Outcome Treatment
1 Unknown