FDA Adverse Event
Injury
Summary report: N
TENNIS ELBOW ECLIPSE
MDR report key: 1546122
·
Received November 20, 2009
Report
- Report Number
- 2020250-2009-00002
- Event Type
- Injury
- Date Received
- November 20, 2009
- Date of Event
- October 5, 2009
- Report Date
- November 19, 2009
- Manufacturer
- OSSUR AMERICAS
- Product Code
- IQI
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUCTIONS FOR USE STATES "DO NOT APPLY HEATED GEL PAD DIRECTLY TO SKIN. IF TOO HOT, ALLOW TO COOL BEFORE APPLYING." THE INSTRUCTIONS FOR USE WILL BE CHANGED FROM 20 TO 10 SECOND HEATING INTERVALS JUST AS A FURTHER PRECAUTION IN ADDITION TO THE OTHER WARNINGS (I.E. NOT WEAR DIRECTLY AGAINST THE SKIN AND TO ALLOW TO COOL BEFORE APPLYING). THIS IFU IS SCHEDULED FOR IMMEDIATE PRINTING.
Description of Event or Problem · 1
PATIENT WARMED UP A TENNIS ELBOW ECLIPSE IN THE MICROWAVE FOR 10 SECONDS AND AFTER WEARING THE PRODUCT NOTED A LARGE RED SPOT AND BLISTER WHERE THE PRODUCT HAD BEEN. THE BLISTER LATER POPPED AND REQUIRED TREATMENT BY IN-HOUSE PHYSICIAN. REP STATED THAT THE PATIENT HAD FULLY RECOVERED FROM THE BLISTER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TENNIS ELBOW ECLIPSE | TENNIS ELBOW BRACE | IQI | OSSUR AMERICAS | E310000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Required Intervention |