FDA Adverse Event Malfunction Summary report: N

BD BBL¿ PLATE TSA5% SB//LEVINE EMB

MDR report key: 15460028 · Received September 20, 2022

Report

Report Number
1119779-2022-01196
Event Type
Malfunction
Date Received
September 20, 2022
Date of Event
September 1, 2022
Report Date
September 30, 2022
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
JSG
PMA / PMN Number
PREAMENDMENT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY: 252258 LOT#2172382 WE CONFIRMED ISSUE AS A REPORT FROM PHOTO. THE ISSUE WAS CONTAMINATION. PHOTO SHOWED THE BACTERIA ON THE BLOOD AGAR. NO ISSUE IN RETENTION SAMPLE. NO ISSUE IN DEVICE HISTORY RECORD. ROOT CAUSE WAS UNDETERMINED. NO TREND. WE WILL CONTINUE TO MONITOR THIS ISSUE.

Additional Manufacturer Narrative · 0

(B)(4). A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD BBL¿ PLATE TSA5% SB//LEVINE EMB HAD FIVE PLATES CONTAMINATED WITH MOLD. THERE WAS NO PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: GET MOULDY.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD BBL¿ PLATE TSA5% SB//LEVINE EMB HAD FIVE PLATES CONTAMINATED WITH MOLD. THERE WAS NO PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: GET MOLDY

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2751545 BD BBL¿ PLATE TSA5% SB//LEVINE EMB CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL JSG BECTON, DICKINSON & CO. (SPARKS) 2172382

Patients

Seq Age Sex Outcome Treatment
1 Unknown