SOUNDSTAR ECO 8F ULTRASOUND CATHETER
Report
- Report Number
- 3023245-2022-00009
- Event Type
- Injury
- Date Received
- September 20, 2022
- Date of Event
- August 24, 2022
- Report Date
- September 20, 2022
- Manufacturer
- BIOSENSE WEBSTER INC
- Product Code
- OBJ
- UDI-DI
- 10846835008852
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Distributor report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE MANUFACTURER PRODUCT ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. MANUFACTURER'S REFERENCE NUMBER: (B)(4). MANUFACTURER'S REPORT # 2029046-2022-02235. IMPORTER'S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED BY MANUFACTURER BIOSENSE WEBSTER INC (BWI) THAT AN UNKNOWN FEMALE PATIENT UNDERWENT AN IDIOPATHIC VENTRICULAR TACHYCARDIA (IDVT) ABLATION PROCEDURE WITH A SOUNDSTAR ECO 8FG ULTRASOUND CATHETER, SOUNDSTAR ECO 8F ULTRASOUND CATHETER AND AN OPTRELL¿ MAPPING CATHETER WITH TRUEREF¿ TECHNOLOGY. THE OPTRELL AND SOUNDSTAR BECAME TANGLED AND THE PATIENT SUFFERED VASCULAR INJURY REQUIRING INTERVENTION. DURING A PVC CASE, THE PATIENT SUFFERED A VASCULAR INJURY. THE BWI COMPANY REPRESENTATIVE STATED THE OPTRELL CATHETER GOT ENTANGLED WITH THE ICE (INTRACARDIAC ECHOCARDIOGRAPHY) CATHETER. THEY WERE MOVING THRU THE ATRIUM AND THE ICE CATHETER WENT THRU ONE OF THE OPTRELL SPLINES. THEY KEPT TRYING TO PULL THE OPTRELL CATHETER AND EVENTUALLY THEY PULLED THE SHEATH OUT THAT THE OPTRELL CATHETER WAS IN. THE PROXIMAL PADDLE THEN GOT ENTRAPPED WITH THE SHAFT OF THE ICE CATHETER. AS THEY CONTINUED TO PULL THE OPTRELL, BOTH THE ICE CATHETER AND THE SHEATH IT WAS IN, WENT OUT OF THE VEIN AND INTO THE TISSUE. THE PATIENT STARTED BLEEDING FROM THE LARGE HOLE. A VASCULAR SURGEON WAS CALLED AND HE TOLD THEM TO PULL IT ALL OUT OF THE SKIN. THEY PULLED THE OPTRELL, THE SHAFT OF THE ICE CATHETER AND THE SHEATH OUT OF THE SKIN AND THEN CUT THE OPTRELL CATHETER AND CUT THE ICE CATHETER AND THEY WERE ABLE TO REMOVE EVERYTHING. IT WAS ALSO NOTED THAT BEFORE THE CATHETER WAS PULLED, FLUOROSCOPY WAS NOT USED TO LOOK AT THE CATHETER. THE FELLOW TOLD THE PHYSICIAN TO LOOK ON FLUORO BEFORE THE CATHETER WAS PULLED BUT THEY DID NOT. THE MEDICAL INTERVENTION PROVIDED WAS A PURSE STITCH TO STOP THE BLEEDING OF THE VEIN AND PRESSURE WAS HELD FOR A LONG TIME. THE PATIENT WAS REPORTED TO BE IN STABLE CONDITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 126092 | SOUNDSTAR ECO 8F ULTRASOUND CATHETER | CATHETER, ULTRASOUND, INTRAVASCULAR | OBJ | BIOSENSE WEBSTER INC | 10439011 | E8396926 | 10846835008852 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Female | Life Threatening| R |