FDA Adverse Event Injury Summary report: N

RX ACCULINK CAROTID STENT SYSTEM

MDR report key: 1545644 · Received November 24, 2009

Report

Report Number
3004742046-2009-00325
Event Type
Injury
Date Received
November 24, 2009
Date of Event
November 5, 2009
Report Date
November 9, 2009
Manufacturer
ABBOTT VASCULAR-VASCULAR SOLUTIONS
Product Code
NIM
PMA / PMN Number
P040038
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

STUDY EVENT. DEVICE #2: RX VIATRAC 14 PLUS DILATATION CATHETER (PART# 1008195-30, LOT# UNK) REFERENCED IS BEING FILED UNDER A SEPARATE MEDWATCH REPORT. EVAL SUMMARY: THE STENT REMAINS IN THE PT. VASOSPASM IS A KNOWN POTENTIAL ADVERSE EVENT ASSOCIATED WITH THE USE OF THE DEVICE AS INDICATED IN THE RX ACCULINK INSTRUCTIONS FOR USE. A REVIEW OF THE DEVICE LOT HISTORY RECORD DID NOT REVEAL ANY NON-CONFORMITIES WHICH COULD HAVE CONTRIBUTED TO THE REPORTED EVENT.

Description of Event or Problem · 1

DEVICE #1 MALFUNCTION: NONE. SYMPTOMS/AE: VASOSPASM. TIME OF SYMPTOMS/AE: POST PROCEDURE. IT WAS REPORTED THAT THE DURING A LEFT INTERNAL CAROTID ARTERY STENTING PROCEDURE, AFTER POST-DILATATION OF THE RX ACCULINK STENT USING THE RX VIATRAC 14 PLUS DILATATION CATHETER, EMBOLI FROM A PRE-EXISTING THROMBUS CREATED SLOW FLOW. AN EMBOLECTOMY WAS PERFORMED USING A NON-ABBOTT ASPIRATION CATHETER AS WELL AS A MULTIPURPOSE CATHETER WITH SUCCESSFUL REMOVAL OF THE CLOT. AFTER REMOVAL OF THE EMBOLIC PROTECTION FILTER, A DISTAL VASOSPASM WAS NOTED. NITROGLYCERIN WAS ADMINISTERED AND THE VASOSPASM RESOLVED. INTRACRANIAL CIRCULATION REMAINED INTACT. THERE WAS NO ADVERSE PATIENT SEQUELAE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RX ACCULINK CAROTID STENT SYSTEM NIM ABBOTT VASCULAR-VASCULAR SOLUTIONS NA 9092451

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention (PART# 1008195-30, LOT# UNK)| BIVALIRUDIN| (PART # 1011649-65, LOT# 9052151)| RX VIATRAC 14 PLUS DILATATION CATHETER:| RX ACCUNET EMBOLIC PROTECTION DEVICE: