FDA Adverse Event Malfunction Summary report: N

CAPTURE-R READY INDICATOR RED CELLS

MDR report key: 1545305 · Received November 25, 2009

Report

Report Number
1034569-2009-00423
Event Type
Malfunction
Date Received
November 25, 2009
Date of Event
October 28, 2009
Report Date
November 25, 2009
Manufacturer
IMMUCOR
Product Code
KSZ
PMA / PMN Number
BK020053
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

REVIEW OF INSTRUMENT RESULTS: CRRID STRIP LOT ID119, CAPTURE-R READY INDICATOR RED CELLS (CRIRC) LOT 221422. 0 STRIP 1 WELL 1=1 (HOMOZYGOUS FOR C ) ( REVIEW OF THE IMAGE APPEARS NEGATIVE); 0 STRIP 1 WELL 2=1 (HOMOZYGOUS FOR C ) ( REVIEW OF THE IMAGE APPEARS NEGATIVE); 0 STRIP 1 WELL 5=1 (HETEROZYGOUS FOR C) ( REVIEW OF THE IMAGE APPEARS NEGATIVE); 0 STRIP 2 WELL 14=1 (HOMOZYGOUS FOR C ) ( REVIEW OF THE IMAGE APPEARS NEGATIVE); ALL OTHER WELLS WERE ANTIGEN NEGATIVE AND REACTED NEGATIVE AS EXPECTED . CRRS3 STRIP LOT R058, CAPTURE-R READY INDICATOR CELLS; 0 STRIP 2 WELL 1=2 (HOMOZYGOUS FOR C ) ( REVIEW OF THE IMAGE APPEARS NEGATIVE); ALL OTHER WELLS WERE ANTIGEN NEGATIVE AND REACTED NEGATIVE AS EXPECTED . CRRID STRIP LOT ID120, CAPTURE-R READY INDICATOR CELLS LOT 221422; 0 STRIP 1 WELL 1=1 HOMOZYGOUS FOR C ) ( REVIEW OF THE IMAGE APPEARS NEGATIVE); 0 STRIP 1 WELL 2=1 HOMOZYGOUS FOR C ) ( REVIEW OF THE IMAGE APPEARS NEGATIVE); 0 STRIP 1 WELL 5=2(HETEROZYGOUS FOR C) ( REVIEW OF THE IMAGE APPEARS NEGATIVE); 0 STRIP 2 WELL 14=2 HOMOZYGOUS FOR C ) ( REVIEW OF THE IMAGE APPEARS NEGATIVE); ALL OTHER WELLS WERE ANTIGEN NEGATIVE AND REACTED NEGATIVE AS EXPECTED . CONFIRMED THE REACTIVITY OF THE C ANTIGEN ON RETENTION CRRID LOT ID120, USING CRIRC LOT 221422. CRRID ASSAY PERFORMED WITH CUSTOMER'S PATIENT SAMPLE (REPORTED TO CONTAIN ANTI-C) USING RETENTION CRRID, LOT ID120, AND CRIRC, LOT 221429, ON IN-HOUSE ECHO. SAMPLE EXHIBITED 2+ TO 3+ POSITIVE REACTIVITY WITH ALL C+ CELLS AND WAS NONREACTIVE WITH ALL C- CELLS. ALL RETENTION PRODUCTS PERFORMED AS EXPECTED.

Description of Event or Problem · 1

CUSTOMER REPORTED A PATIENT SAMPLE WITH KNOWN HISTORY OF ANTI-C WAS TESTED ON THE ECHO 3 TIMES, FIRST WITH CAPTURE-R READY ID (CRRID) LOT ID 119, SECOND WITH CAPTURE-R READY SCREEN 3 (CRRS 3) AND A THIRD TIME WITH CRRID LOT ID120. THE RESULTS WERE NEGATIVE WITH ALL THREE ASSAYS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPTURE-R READY INDICATOR RED CELLS ANTI-IGG COATED RED CELLS FOR USE IN SOLID PHASE ASSAY KSZ IMMUCOR 221422

Patients

Seq Age Sex Outcome Treatment
1 35 YR