GN RESOUND
Report
- Report Number
- 3005650109-2022-00035
- Event Type
- Malfunction
- Date Received
- September 20, 2022
- Date of Event
- August 22, 2022
- Report Date
- October 18, 2022
- Manufacturer
- GN HEARING A/S
- Product Code
- OSM
- UDI-DI
- 05704819289933
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- 003
Narratives
SF: (B)(4). A DHR REVIEW HAVE BEEN COMPLETED. NO DEVIATION OR CHANGES WERE FOUND DURING MANUFACTURING OF THE DEVICE. IT HAS BEEN REPORTED- NOT ENOUGH INFORMATION HOW THE HA GOT MELTED IN THE DRY JAR. AID PHYSICALLY MELTED AND BROKEN INTO PIECES. TECHNICAL INVESTIGATION: MOST DRYING JARS CAN DRY AIDS IN SURROUNDINGS UP TO 50-60 DEGREES. IF THE HEARING AIDS ARE TURNED OFF, IT IS FINE TO USE DRY JARS. A DRY BOX CAN BE USED FOR ALL TYPES OF GN HEARING AIDS.IT IS SUSPECTED THAT THE AIDS WERE LEFT IN THE DRY BOX SUBJECTED TO SURROUNDINGS WITH EXCESSIVE HEAT. THIS IS AN INITIAL REPORT.
(B)(6). A DHR REVIEW HAVE BEEN COMPLETED. NO DEVIATION OR CHANGES WERE FOUND DURING MANUFACTURING OF THE DEVICE. IT HAS BEEN REPORTED- NOT ENOUGH INFORMATION HOW THE HA GOT MELTED IN THE DRY JAR. AID PHYSICALLY MELTED AND BROKEN INTO PIECES. TECHNICAL INVESTIGATION: MOST DRYING JARS CAN DRY AIDS IN SURROUNDINGS UP TO 50-60 DEGREES. IF THE HEARING AIDS ARE TURNED OFF, IT IS FINE TO USE DRY JARS. A DRY BOX CAN BE USED FOR ALL TYPES OF GN HEARING AIDS.IT IS SUSPECTED THAT THE AIDS WERE LEFT IN THE DRY BOX SUBJECTED TO SURROUNDINGS WITH EXCESSIVE HEAT. CLINICAL EVALUATION OF THE EVENT: CASE INFORMATION STATES THAT THE DEVICES MELTED WHILE PLACED IN A DRYING JAR. THE JAR IS NOT A GN HEARING PRODUCT, AND THE PATIENT REPORTS THAT THE HEATING ELEMENT IN THE DRYING JAR MAY HAVE CAUSED THE MELTING. NO PATIENT HARM WAS REPORTED, AS THE DEVICES WERE IN THE DRYING JAR AND NOT IN THE PATIENT'S EARS. DEVICE USER GUIDE HAS A WARNING HOW TO DRY THE DEVICES. RISK ANALYSIS ALREADY INCLUDED, THE RESIDUAL RISK AFTER THE RISK CONTROL MEASURE(S) IS ACCEPTABLE. THIS IS A FINAL REPORT. NO FURTHER INFORMATION IS EXPECTED.
SF: (B)(4). IT WAS REPORTED- AIDS PUT IN DRY JAR AND MELTED. CUSTOMER VERIFIED THE JAR WAS NOT RESOUND BRAND. THEY BELIEVE THE HEAT ELEMENT IN THE JAR CAUSED THE MELTING AND THEREFORE WAS NOT RELATED TO THE RESOUND AIDS, BUT NO SPECIFIC INVESTIGATION WAS DONE' THE CURRENT INFORMATION INDICATES NEED FOR IMMEDIATE REPORTING. BASED ON THE AVAILABLE INFORMATION, THE EVENT/ INCIDENT DID LEAD TO, OR MIGHT LED TO, DEATH OR A SERIOUS INJURY IF IT WERE TO RECUR. THIS IS AN INITIAL REPORT.
(B)(6). IT WAS REPORTED- AIDS PUT IN DRY JAR AND MELTED. CUSTOMER VERIFIED THE JAR WAS NOT RESOUND BRAND. THEY BELIEVE THE HEAT ELEMENT IN THE JAR CAUSED THE MELTING AND THEREFORE WAS NOT RELATED TO THE RESOUND AIDS, BUT NO SPECIFIC INVESTIGATION WAS DONE'. THE CURRENT INFORMATION INDICATES NEED FOR REPORTING. BASED ON THE AVAILABLE INFORMATION, THE EVENT/ INCIDENT DID LEAD TO, OR MIGHT LED TO, DEATH OR A SERIOUS INJURY IF IT WERE TO RECUR. THIS IS A FINAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2698693 | GN RESOUND | HEARING AID, AIR CONDUCTION WITH WIRELESS TECHNOLOGY | OSM | GN HEARING A/S | 19706708 | 05704819289933 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |