FDA Adverse Event Malfunction Summary report: N

BD VEO¿ INSULIN SYRINGES WITH BD ULTRA-FINE¿ NEEDLE

MDR report key: 15452515 · Received September 19, 2022

Report

Report Number
1920898-2022-00629
Event Type
Malfunction
Date Received
September 19, 2022
Date of Event
August 24, 2022
Report Date
November 4, 2022
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
K190054
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. DEVICE EVALUATED BY MFR: A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY CUSTOMER RETURNED (1) LOOSE 0.5ML BD INSULIN SYRINGE. THE CUSTOMER REPORTED THAT THE NEEDLE HUB SEPARATED. THE SAMPLE WAS EXAMINED, AND IT WAS OBSERVED THAT THE NEEDLE HUB/SHIELD ASSEMBLY WAS DETACHED FROM THE BARREL. NO DAMAGE TO THE BARREL TIP WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 1130878. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. EMBECTA WAS ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE. A CAPA HAS BEEN OPENED TO ADDRESS THIS ISSUE. H3 OTHER TEXT : SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD VEO¿ INSULIN SYRINGES WITH BD ULTRA-FINE¿ NEEDLE SEPARATED AND STAYED INSIDE OF SHIELD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CONSUMER REPORTED THAT THE NEEDLE HUB SEPARATED AND STAYED INSIDE OF THE SHIELD.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD VEO¿ INSULIN SYRINGES WITH BD ULTRA-FINE¿ NEEDLE SEPARATED AND STAYED INSIDE OF SHIELD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CONSUMER REPORTED THAT THE NEEDLE HUB SEPARATED AND STAYED INSIDE OF THE SHIELD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2694560 BD VEO¿ INSULIN SYRINGES WITH BD ULTRA-FINE¿ NEEDLE PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 1130878

Patients

Seq Age Sex Outcome Treatment
1 Unknown