BD ULTRA-FINE¿ II INSULIN SYRINGE
Report
- Report Number
- 1920898-2022-00625
- Event Type
- Malfunction
- Date Received
- September 19, 2022
- Date of Event
- August 23, 2022
- Report Date
- November 1, 2022
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- 003
Narratives
DEVICE EVALUATED BY MFR: A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY: CUSTOMER RETURNED (7) 3/10CC, 8MM, 30G SYRINGES IN AN OPEN POLY BAG FROM LOT # 1256984. CUSTOMER STATES THAT DARK DIRT WAS FOUND ONTO THE BARREL. ALL RETURNED SYRINGES WERE EXAMINED AND 3 OUT OF 7 SAMPLES EXHIBITED INK SMEARED ON THE SURFACE OF THE BARREL. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 1256984. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. EMBECTA WAS ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE. ROOT-CAUSE: DISPATCH 133669: DIRTY SPINDLE CUP; DOCTOR BLADE WAS NOT SEATED ALL THE WAY DOWN THE SHAFT CAUSING INK TO FILL THE SPINDLE CUPS AND COAT THE PADS. H3 OTHER TEXT : SEE H10.
IT WAS REPORTED THAT DIRT-LIKE FOREIGN MATTER WAS FOUND IN THE BARRELS OF 2 BD ULTRA-FINE¿ II INSULIN SYRINGES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE: THE USER REPORTED THAT THE DARK DIRT WAS FOUND ONTO THE BARREL.
IT WAS REPORTED THAT DIRT-LIKE FOREIGN MATTER WAS FOUND IN THE BARRELS OF 2 BD ULTRA-FINE¿ II INSULIN SYRINGES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE: "THE USER REPORTED THAT THE DARK DIRT WAS FOUND ONTO THE BARREL."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2698659 | BD ULTRA-FINE¿ II INSULIN SYRINGE | PISTON SYRINGE | FMF | BD MEDICAL - DIABETES CARE | 1256984 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |