FDA Adverse Event Injury Summary report: N

STYLE 15 SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 15451240 · Received September 19, 2022

Report

Report Number
9617229-2022-15831
Event Type
Injury
Date Received
September 19, 2022
Date of Event
August 15, 2022
Report Date
November 15, 2022
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
UDI-DI
10888628001473
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: BASED ON THE DEVICE ANALYSIS GRID, THE ASSESSMENTS OF THE COMPLAINT ARE: RUPTURE: OBSERVED, ONE OPENING IN THE RADIUS SIDE ASSESSED AS SURGICAL DAMAGE AND MISSING SIDE ASSESSED AS INCONCLUSIVE . NO OTHER ADDITIONAL OBSERVATION.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE. DEVICE REMAINS IMPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL ADDITIONALLY REPORTED FREE SILICONE IN THE POCKET VIA AN RGA FORM. DEVICE HAS BEEN EXPLANTED AND REPLACED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED RIGHT-SIDE RUPTURE AND "FREE SILICONE IN THE POCKET" VIA AN RGA FORM. THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
126942 STYLE 15 SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 15-371 2697704 10888628001473

Patients

Seq Age Sex Outcome Treatment
1 44 YR Female Required Intervention