IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM
Report
- Report Number
- 0002023141-2022-02423
- Event Type
- Injury
- Date Received
- September 19, 2022
- Date of Event
- July 13, 2022
- Report Date
- February 28, 2023
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024019508
- PMA / PMN Number
- K061410
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER (B)(4). ADDITIONAL PMA/510(K) NUMBERS ARE K011028 AND K013227.
THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: B4: DATE OF THIS REPORT WAS UPDATED. G3: DATE RECEIVED BY MANUFACTURER WAS UPDATED. G6: TYPE OF REPORT WAS UPDATED. H2: TYPE OF FOLLOW UP WAS UPDATED. H3: DEVICE EVALUATED BY MANUFACTURER WAS UPDATED. H6: TYPE OF INVESTIGATION CODES WERE UPDATED: 3331, 4109, 4110, 4111, 4114. H6: INVESTIGATION FINDINGS CODE WAS UPDATED: 213. H6: INVESTIGATION CONCLUSION CODE WAS UPDATED: 4315. H10: NARRATIVE/DATA WAS UPDATED. ONE (1) IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM (TSVB10) WAS NOT RETURNED FOR INVESTIGATION. VISUAL INSPECTION COULD NOT BE PERFORMED. THE INVESTIGATION HAS BEEN PERFORMED BASED ON THE AVAILABLE INFORMATION USING THE ITEM AND LOT NUMBER ALONG WITH THE APPLICABLE DEVICE HISTORY RECORD (DHR), INSTRUCTIONS FOR USE (IFU), AND RISK FILE. BASED ON THE EVALUATION, THE DEVICE MALFUNCTION COULD NOT BE VERIFIED. ADDITIONALLY, THERE IS NO EXISTING NONCONFORMANCE/CAPA/HHE/D/IE/PRODUCT HOLDS AGAINST THE REPORTED DEVICE THAT DID OR COULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. MONTHLY POST MARKET TRENDING REVIEW IDENTIFIED NO ACTIONABLE TRENDS OR CORRECTIVE ACTIONS FOR THE REPORTED EVENT OR DEVICE. ZIMMER BIOMET QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO ENSURE THE DISTRIBUTION OF CONFORMING PRODUCT. THEREFORE, BASED ON THE AVAILABLE INFORMATION, THE PRODUCT WAS WITHIN SPECIFICATIONS AND LIKELY CONFORMING WHEN IT LEFT ZIMMER BIOMET. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1244047). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1244047) FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. THE COMPLAINT IS RELATED TO THE FUNCTIONAL PERFORMANCE OF THE DEVICE. A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. HOWEVER, BASED ON THE INVESTIGATION AND RISK FILE REVIEW, THE MOST LIKELY CAUSE DETERMINED FROM THE INVESTIGATION IS CLINICIAN ERROR -- IMPROPER SURGICAL PROCEDURE. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA/HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.
THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. ADDITIONAL INFORMATION: A REPLACEMENT IMPLANT WAS UNABLE TO BE PLACED. THE FOLLOWING SECTIONS ARE BEING REPORTED: B1: REPORT TYPE WAS UPDATED. B2: OUTCOMES ATTRIBUTED TO ADVERSE EVENT WAS UPDATED. B4: DATE OF THIS REPORT WAS UPDATED. G3: DATE RECEIVED BY MANUFACTURER WAS UPDATED. G6: TYPE OF REPORT WAS UPDATED. H1: TYPE OF REPORTABLE EVENT WAS UPDATED. H2: TYPE OF FOLLOW UP WAS UPDATED. H10: NARRATIVE/DATA WAS UPDATED.
IT WAS REPORTED THAT THE MOUNT WOULD NOT DISENGAGE FROM THE IMPLANT AT THE TIME OF PLACEMENT.
NO ADDITIONAL EVENT INFORMATION RECEIVED AT THIS TIME OF THIS REPORT.
UPON FOLLOW UP, THE IMPLANT COORDINATOR STATED THAT THEY DID NOT COMPLETE THE PROCEDURE WITH ANOTHER DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2837378 | IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | TSVB10 | 1244047 | 00889024019508 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Male | Required Intervention |