NEURON DELIVERY CATHETER 070 ADVANCED
Report
- Report Number
- 3005168196-2009-00095
- Event Type
- Malfunction
- Date Received
- November 9, 2009
- Date of Event
- October 15, 2009
- Report Date
- October 15, 2009
- Manufacturer
- PENUMBRA, INC.
- Product Code
- DQY
- PMA / PMN Number
- K082290
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
Narratives
TECHNICAL EVAL: THE FIRST 0.5 CM OF THE DISTAL TIP WAS COMPLETELY FLATTENED AND THERE WERE SINGLE AXIS BENDS AT 1.8 AND 2.5CM. THERE WAS ALSO A LONGER FLATTENED SPOT BETWEEN 4.5 AND 6.2CM FROM THE TIP. THE INCIDENT IS CONFIRMED AS REPORTED. IT SHOULD BE NOTED THAT BOTH THE TIP FLATTENING AND THE SINGLE AXIS BENDS ARE UNUSUAL FOR THIS SORT OF FAILURE AND ARE LIKELY THE RESULT OF INATTENTIVE HANDLING AFTER THE INCIDENT. FLATTENED CATHETER TIPS ARE REPORTABLE INCIDENTS PER FDA GUIDANCE AS OF 28 AUG 2009 (INCLUDING OUT-OF-BOX FAILURES). A DHR OF THIS MFG LOT HAS BEEN REVIEWED. QTY: 1. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS COMPLETED ON PART # 2314-03, (LOT # L15514). ALL APPROPRIATE INSPECTIONS WERE COMPLETED PER PROCESS MONITORING REQUIREMENTS OR 100% INSPECTIONS WHERE REQUIRED. IN ADDITION, ALL DESTRUCTIVE TESTING WAS COMPLETED PER REQUIRED PROCESS MONITORING REQUIREMENTS AND RESULTS MET SPEC FOR THE TENSILE STRENGTH. ALL PARTS THAT FAILED VISUAL INSPECTION HAVE BEEN REJECTED AND ALL PARTS MEETING SPECS WERE RELEASED FOR SHIPMENT. LOT# L15514 WAS ASSOCIATED WITH (B)(4) DUE TO "SHORT-SHOTS" IN THE HUB MOLDING PROCESS. THIS (B)(4) HAS NOT BEEN FOUND TO BE RELATED TO THIS INCIDENT WHICH INVOLVED THE NEURON TIP, NOT THE HUB. THE PARTS RELEASED IN THESE LOTS MET PROCESS REQUIREMENTS.
THE TIP OF THE NEURON 070 WAS FOUND TO BE FLATTENED UPON OPENING THE PACKAGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEURON DELIVERY CATHETER 070 ADVANCED | SUBCUTANEOUS CATHETER | DQY | PENUMBRA, INC. | F15514 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |