FDA Adverse Event Malfunction Summary report: N

MYA JOY

MDR report key: 15428059 · Received September 15, 2022

Report

Report Number
3009974348-2022-00356
Event Type
Malfunction
Date Received
September 15, 2022
Date of Event
August 18, 2022
Report Date
August 19, 2022
Manufacturer
AMEDA, INC.
Product Code
HGX
PMA / PMN Number
K172989
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

RETURNED MYA JOY AC ADAPTER WAS NOT TESTED CITING SAFETY CONCERNS. DAMAGE TO THE 120-VOLT TRANSFORMER HOUSING, PRONGS AND THEIR CONNECTION TO THE PCB INDICATES THAT THE RETURNED AC ADAPTER MOST LIKELY SUSTAINED DAMAGE PRIOR TO SPARK EVENT. AMEDA WAS UNABLE TO DETERMINE THE FAULT OF THE 120-VOLT TRANSFORMER HOUSING AND MISSING PRONG DAMAGE.

Description of Event or Problem · 0

CUSTOMER CONTACTED AMEDA, INC. ON (B)(6) 2022 TO REPORT SPARKING FROM THE MYA JOY POWER ADAPTER AS SHE PLUGGED IT INTO AN OUTLET AT WORK ON (B)(6) 2022. SHE IMMEDIATELY PULLED OUT THE ADAPTER FROM THE OUTLET BY GRASPING THE BLACK PLASTIC HOUSING WHICH THEN BROKE APART. ONE OF THE TWO PRONGS FELL OUT. CUSTOMER DENIED INJURY TO HERSELF. CUSTOMER WAS OVERNIGHT SHIPPED A REPLACEMENT MYA JOY POWER ADAPTER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
271016 MYA JOY ELECTRIC BREAST PUMP HGX AMEDA, INC. LD-306 0920

Patients

Seq Age Sex Outcome Treatment
1 37 YR Female