FDA Adverse Event Malfunction Summary report: N

PULSE GEN MODEL 101

MDR report key: 1542723 · Received November 17, 2009

Report

Report Number
1644487-2009-02572
Event Type
Malfunction
Date Received
November 17, 2009
Date of Event
May 7, 2002
Report Date
November 11, 2009
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

A VNS PT HAD GENERATOR REPLACEMENT SURGERY DUE TO END OF SERVICE. THE EXPLANTED GENERATOR WAS RETURNED FOR PRODUCT ANALYSIS AND IT WAS IDENTIFIED THAT TRANSISTOR Q10 WAS FOUND TO EXHIBIT AN OUT-OF-LIMIT (HIGHER) LEAKAGE CURRENT AT TEST CHAMBER TEMPERATURE. THIS LEAKAGE INCREASED THE MODULE'S SUPPLY CURRENT 1MA CONSUMPTION BY APPROX 1.194UA OVER THE HIGH LIMIT (SPEC 23.000-38.000MA). THE OUT-OF-LIMIT PULSING CURRENTS COULD POTENTIALLY BE A CONTRIBUTING FACTOR TO THE END OF SERVICE, HOWEVER, THE BATTERY LONGEVITY CALCULATION (-0.33 YEARS) DETERMINED THAT THE END OF SERVICE WAS AN EXPECTED EVENT. AS THE GENERATOR WAS RECEIVED IN AN END OF SERVICE STATE, THE EXTENT TO WHICH THE INCREASED CURRENT CONSUMPTION MAY HAVE CONTRIBUTED TO A DEPLETED BATTERY CANNOT BE DETERMINED. THE CAGED MODULE MET SPEC FOLLOWING THE Q10 REPLACEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PULSE GEN MODEL 101 LYJ CYBERONICS, INC. 101 5985

Patients

Seq Age Sex Outcome Treatment
1 22 YR